Image unavailable
Image unavailable
Rosemount
Rosemount Xi Advanced Electronics
السعر عند الطلب
بدون دفع عبر الإنترنت · أضف بنودًا إلى قائمة طلب العرض
تنزيل ورقة المواصفات PDF
طلب العرض يحمل بنود قائمة المواد فقط. أرسل لعرض هندسي.
المعاملات الرئيسية
Measuring range
Advanced electronics for ZrO2 flue-gas O2 probes (Oxymitter 4000 / World Class / most Rosemount probes); probe O2 accuracy ±0.75% of reading or ±0.05% O2
Accuracy
Outstanding accuracy with probe; AMS-aware HART diagnostics including calibration-recommended alarm
Output
HART® communications standard; 4–20 mA; optional WirelessHART via THUM™ Adaptor
Supply
100–240 V ±10%, 50–60 Hz (catalog); Xi power consumption 10 W / 12 VA max (traditional architecture higher with probe heater)
Ambient temperature
−20 to +50 °C (−4 to +122 °F) Xi ambient; −20 to +55 °C inside electronics housing / LCD limits
Display
Large backlit LCD local operator interface
| المعامل | القيمة |
|---|---|
| Measuring range | Advanced electronics for ZrO2 flue-gas O2 probes (Oxymitter 4000 / World Class / most Rosemount probes); probe O2 accuracy ±0.75% of reading or ±0.05% O2 |
| Accuracy | Outstanding accuracy with probe; AMS-aware HART diagnostics including calibration-recommended alarm |
| Output | HART® communications standard; 4–20 mA; optional WirelessHART via THUM™ Adaptor |
| Supply | 100–240 V ±10%, 50–60 Hz (catalog); Xi power consumption 10 W / 12 VA max (traditional architecture higher with probe heater) |
| Ambient temperature | −20 to +50 °C (−4 to +122 °F) Xi ambient; −20 to +55 °C inside electronics housing / LCD limits |
| Display | Large backlit LCD local operator interface |
| Approvals | CSA, FM & CE (catalog); Xi electronics general-purpose / safe area only |
| Mounting | None / panel mount kit with gasket / 2 in pipe-wall mount kit |
| Sensor type | Remote advanced electronics for zirconium-oxide O2 probes; traditional-architecture and probe-mounted LOI options |
| Notes | Specs from Emerson catalog Rosemount Xi Advanced Electronics + Oxymitter 4000 PDS 00813-0100-4340 (Xi ordering section). Distinct from 6888 Xi (1207-0002). |
| Special Features | Fully automatic calibration with solenoid box; extended process temperature; programmable reference; stoichiometer; works with Westinghouse/Rosemount and most ZrO2 probes |
| Alarm Relays | 2 Form-C relays, 2 A @ 30 Vdc |