Rosemount

Rosemount Model 228 Conductivity Sensor

SKU · 36300 · Model 228 Conductivity Sensor · العلامة: Rosemount
متوفر
السعر عند الطلب
بدون دفع عبر الإنترنت · أضف بنودًا إلى قائمة طلب العرض
تنزيل ورقة المواصفات PDF

طلب العرض يحمل بنود قائمة المواد فقط. أرسل لعرض هندسي.

المعاملات الرئيسية

Measuring range
Toroidal conductivity up to 2 S/cm (2,000,000 µS/cm) in highly conductive / dirty / corrosive solutions
Accuracy
Nominal cell constant 2.7/cm (PDS lists ~3/cm); field calibration required; wall-effect clearance ≥1 in from metal walls
Process connection
⅝-11 UNC (EPDM gasket; requires adapter) or ¾ in MNPT
Wetted parts
Glass-filled PEEK, glass-filled Tefzel®, or unfilled Tefzel body options
Sensor type
Toroidal (inductive) insertion/submersion conductivity sensor with integral PT-100 RTD
Notes
Specs from Emerson Rosemount 228 PDS 00813-0100-3228 Rev AD + QSG 00825-0100-3228.
تنزيل ورقة المواصفات PDF
المعاملالقيمة
Measuring rangeToroidal conductivity up to 2 S/cm (2,000,000 µS/cm) in highly conductive / dirty / corrosive solutions
AccuracyNominal cell constant 2.7/cm (PDS lists ~3/cm); field calibration required; wall-effect clearance ≥1 in from metal walls
Process connection⅝-11 UNC (EPDM gasket; requires adapter) or ¾ in MNPT
Wetted partsGlass-filled PEEK, glass-filled Tefzel®, or unfilled Tefzel body options
Sensor typeToroidal (inductive) insertion/submersion conductivity sensor with integral PT-100 RTD
NotesSpecs from Emerson Rosemount 228 PDS 00813-0100-3228 Rev AD + QSG 00825-0100-3228.
Process TemperatureStandard up to 120 °C (248 °F); high-temp PEEK option up to 200 °C (392 °F)
Process PressureUp to 295 psig (21.4 bar) on high-temperature option
CableIntegral cable options 10–100 ft; unshielded (54) or shielded EMI/RFI (56); 20 ft (6.1 m) common
Special FeaturesReinforced internal metal frame for vibration; optional retraction assembly for hot insertion/removal
Compatible TransmittersRecommended with 1056 / 1066-T / 56 / 5081 (shielded cable); also 1054/2054 with unshielded