Rosemount

Rosemount Model T1056 Clarity II Turbidimeter

SKU · 31292 · Model T1056 Clarity II Turbidimeter · العلامة: Rosemount
متوفر
السعر عند الطلب
بدون دفع عبر الإنترنت · أضف بنودًا إلى قائمة طلب العرض
تنزيل ورقة المواصفات PDF

طلب العرض يحمل بنود قائمة المواد فقط. أرسل لعرض هندسي.

المعاملات الرئيسية

Accuracy
After cal at 20.0 NTU: 0–1 NTU ±2% of reading or ±0.015 NTU (whichever greater); 0–20 NTU ±2% of reading
Wetted parts
Delrin, glass, EPDM (sensor/chamber)
Ambient temperature
32 to 131 °F (0 to 55 °C); turbidity-only 32 to 122 °F (0 to 50 °C); RH 5–95% non-condensing
Measurement Range
0 to 200 NTU class (display resolution 0.001 NTU; accuracy stated after 20.0 NTU calibration for 0–1 and 0–20 NTU bands)
Resolution
0.001 NTU (four-digit display; decimal shifts across 0.001–1000 class ranges)
Sensor Methods
US EPA 180.1 (incandescent) or ISO 7027 (860 nm LED); specified at order
تنزيل ورقة المواصفات PDF
المعاملالقيمة
AccuracyAfter cal at 20.0 NTU: 0–1 NTU ±2% of reading or ±0.015 NTU (whichever greater); 0–20 NTU ±2% of reading
Wetted partsDelrin, glass, EPDM (sensor/chamber)
Ambient temperature32 to 131 °F (0 to 55 °C); turbidity-only 32 to 122 °F (0 to 50 °C); RH 5–95% non-condensing
Measurement Range0 to 200 NTU class (display resolution 0.001 NTU; accuracy stated after 20.0 NTU calibration for 0–1 and 0–20 NTU bands)
Resolution0.001 NTU (four-digit display; decimal shifts across 0.001–1000 class ranges)
Sensor MethodsUS EPA 180.1 (incandescent) or ISO 7027 (860 nm LED); specified at order
OutputsTwo isolated 0–20/4–20 mA (max load 550 Ω); optional HART on Output 1; four Form-C SPDT alarm relays
EnclosurePolycarbonate Type 4X/CSA 4 (IP65); overall 6.10×6.10×5.5 in (155×155×131 mm); ½ DIN cutout
Power20–30 Vdc (−02) or 85–265 Vac 47.5–65 Hz (−03); 15 W min
Sensor Cable20 ft (6.1 m) or 50 ft (15.2 m) max; IP65 connector when mated
Sample Temperature40 to 95 °F (5 to 35 °C); max pressure 30 psig (308 kPa)
System ContentsSingle/dual-input 1056 analyzer + sensor(s) + debubbler/measuring chamber + cable per sensor
DocumentPDS 00813-0100-3560 Rev AC May 2024