Time Electronics

Time Electronics 1040 Resistance Decade Box

SKU · 25562 · 1040 · Brand: Time Electronics
Available
Quote on request
No online payment · add lines to RFQ BOM

8-decade resistance box; 1 Ω–100 MΩ; to ±0.1%. RFQ + selection help.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Accuracy
±1% (1 Ω, 10 MΩ); ±0.5% (10 Ω); ±0.1% (100 Ω…1 MΩ mid decades)
Type
8-decade precision resistance decade box
Residual Resistance
Less than 250 mΩ
Power Rating
1 W per resistor
Voltage Rating
Maximum 250 V DC / 176 V AC RMS
Max Current
0.5 A (1 Ω decade) down to µA on high decades
Download datasheet PDF

Fits wide-range resistance substitution where you need a Time Electronics 1040 covering 1 Ω to 100 MΩ in 1 Ω steps with mid-decade accuracy to ±0.1%.

Key advantages

  • Range — 1 Ω to 100 MΩ in 1 Ω steps (8 decades)
  • Accuracy — ±0.1% mid decades; ±0.5%/±1% on extremes
  • Residual — <250 mΩ
  • Ratings — 1 W per resistor; max 250 V DC / 176 V AC RMS
  • TC — 50 ppm/°C; safety terminals

Typical applications

  • Meter and bridge calibration aids.
  • Circuit resistance substitution.
  • Wide-range lab decade resistance.

Selection guide

  • High end — 100 MΩ need
  • Accuracy decade — Mid vs end decades
  • Power/voltage — 1 W / 250 V limits
  • Residual R — <250 mΩ context

Common mistakes

  • Exceeding wattage on low decades.
  • Ignoring residual on low settings.
  • Overvoltage on high decades.

When to RFQ / what to tell us

  1. Quantity; ship-to; calibration cert if needed.

Request a Quote · Get selection help

Specs

SpecValue
SKU25562
BrandTime Electronics
Model1040 Resistance Decade Box
Product typeResistance Decade Box
Type8-decade precision resistance decade box
Range1 Ω to 100 MΩ in 1 Ω steps
Accuracy±1% (1 Ω, 10 MΩ); ±0.5% (10 Ω); ±0.1% (100 Ω…1 MΩ mid decades)
Residual resistanceLess than 250 mΩ
Power rating1 W per resistor
Voltage ratingMaximum 250 V DC / 176 V AC RMS
Max current0.5 A (1 Ω decade) down to µA on high decades
Temperature coefficient50 ppm/°C
TerminalsSafety terminals compatible with 4 mm shrouded plugs / wires / spades

Documents

FAQ

Q: Range? A: 1 Ω to 100 MΩ in 1 Ω steps.

Q: Best accuracy? A: ±0.1% on 100 Ω…1 MΩ mid decades.

Q: What should I send for a quote? A: Quantity and ship-to.

Links

ParameterValue
Accuracy±1% (1 Ω, 10 MΩ); ±0.5% (10 Ω); ±0.1% (100 Ω…1 MΩ mid decades)
Type8-decade precision resistance decade box
Residual ResistanceLess than 250 mΩ
Power Rating1 W per resistor
Voltage RatingMaximum 250 V DC / 176 V AC RMS
Max Current0.5 A (1 Ω decade) down to µA on high decades
Temperature Coefficient50 ppm/°C
TerminalsSafety terminals compatible with 4 mm shrouded plugs / wires / spades
ConstructionMetal case with removable protective rubber cover; gold multi-contact switches
DimensionsW 215 × H 100 × D 120 mm / approx. 1 kg
Document1040 Data Sheet + 1040 User Manual