Fluke 714B Temperature Calibrator
Thermocouple calibrator: 17 TC types + mV, isolated mA measure. IP52 handheld. RFQ + selection help.
RFQ cart holds BOM lines only. Submit for engineering quote.
Key parameters
Fits field thermocouple calibration where you need to source/measure 17 TC types plus mV, and simultaneously measure an isolated mA loop — not a full multi-function documenting calibrator.
Key advantages
- 17 TC types + mV — source/measure thermocouples and −10…75 mV on one tool.
- Isolated mA measure — simultaneous 0–24 mA channel; mA accuracy 0.010% + 2 µA.
- IP52 handheld — 4× AA; magnetic hanger; dual backlit inputs.
- Published TC tables — 1-year and 2-year TC accuracy tables in OEM documentation.
Typical applications
- Field thermocouple calibration of transmitters and recorders.
- Loop checks needing TC source plus isolated mA measure.
- Companion use with RTD-focused 712B on the same cart.
Selection guide
- Measuring / source range — 17 thermocouple types + mV (−10…75 mV); isolated mA measure 0–24 mA.
- Accuracy — mA 0.010% + 2 µA; mV 0.015% + 10 µA (1-year, 18–28 °C); TC accuracies per OEM documentation tables. 1-year and 2-year specifications published.
- Output / channels — TC / mV source; simultaneous mA measure channel. Dual inputs, backlit display.
- Supply — 4× AA NEDA 1.5A IEC LR6.
- Electrical connection — TC mini; banana for mA.
- IP — IP52 (IEC 60529).
- Ambient — Operating −10…+50 °C; storage −30…+60 °C; operating altitude 2000 m.
- Mounting — Handheld; magnetic hanger included.
- Weight / size — 515 g; 52.5 × 84 × 188.5 mm (H×W×L).
- Approvals — IEC 61010-1; IEC 61326-1 Portable; MIL-T-28800E Class 2 vibration.
Common mistakes
- Expecting RTD source/measure — 714B is thermocouple / mV focused with isolated mA measure; confirm RTD need against a different calibrator family.
- Using 1-year accuracy figures outside 18–28 °C without checking the 2-year / extended ambient tables.
- Assuming the mA channel sources loop current — datasheet lists isolated mA measure (0–24 mA), not mA source.
- Ignoring IP52 limits in wet or dusty plant areas beyond that rating.
- Ordering without confirming TC mini vs banana lead set for your sensor connectors.
- Calibration instruments · Calibration buying guide
When to RFQ / what to tell us
- TC types you source/measure most often, and whether mV mode is required.
- Whether simultaneous mA measure (0–24 mA) is needed on the same job.
- Accuracy window (1-year vs 2-year) and ambient you calibrate in.
- Lead/connector needs (TC mini, banana) and magnetic hanger use.
- Quantity, ship-to country, and any documentation beyond the included traceable certificate.
Request a Quote · Get selection help
Specs
| Spec | Value |
|---|---|
| SKU | 40222 |
| Brand | Fluke |
| Model | 714B |
| Product type | calibrator (thermocouple temperature calibrator) |
| Measuring range | 17 thermocouple types + mV (−10…75 mV); isolated mA measure 0–24 mA |
| Accuracy | mA 0.010% + 2 µA; mV 0.015% + 10 µA (1-year, 18–28 °C); TC accuracies per OEM documentation tables |
| Output signal | TC / mV source; simultaneous mA measure channel |
| Supply voltage | 4× AA NEDA 1.5A IEC LR6 |
| Electrical connection | TC mini; banana for mA |
| IP rating | IP52 (IEC 60529) |
| Ambient temperature | Operating −10…+50 °C; storage −30…+60 °C; operating altitude 2000 m |
| Display | Dual inputs, backlit display |
| Mounting | Handheld; magnetic hanger included |
| Weight | 515 g |
| Size | 52.5 × 84 × 188.5 mm (H×W×L) |
| Approvals | IEC 61010-1; IEC 61326-1 Portable; MIL-T-28800E Class 2 vibration |
| Notes | Traceable certificate of calibration included per OEM. 1-year and 2-year specifications published. |
Documents
FAQ
Q: Which sensors does 714B cover? A: 17 thermocouple types plus mV (−10…75 mV), with an isolated mA measure channel 0–24 mA. It is not positioned as an RTD calibrator in this OEM extract.
Q: What accuracy should I quote? A: mA 0.010% + 2 µA and mV 0.015% + 10 µA at 1-year, 18–28 °C; TC tables are in the datasheet. State which TC types and ambient you need.
Q: Is a calibration certificate included? A: OEM states a traceable certificate of calibration is included. Confirm any extra accreditation marks in the RFQ.
Q: What should I send for a quote? A: TC types, mA-measure need, accuracy window, quantity, ship-to. Compatible with workflows using Fluke 714B — we do not claim authorized-dealer status.
Links
| Parameter | Value |
|---|---|
| Measuring range | 17 thermocouple types + mV (−10…75 mV); isolated mA measure 0–24 mA |
| Accuracy | mA 0.010% + 2 µA; mV 0.015% + 10 µA (1-year, 18–28 °C); TC accuracies per datasheet tables |
| Output | TC / mV source; simultaneous mA measure channel |
| Supply | 4× AA NEDA 1.5A IEC LR6 |
| Electrical connection | TC mini; banana for mA |
| Protection | IP52 (IEC 60529) |
| Ambient temperature | Operating −10…+50 °C; storage −30…+60 °C; operating altitude 2000 m |
| Display | Dual inputs, backlit display |
| Approvals | IEC 61010-1; IEC 61326-1 Portable; MIL-T-28800E Class 2 vibration |
| Mounting | Handheld; magnetic hanger included |
| Notes | Size (H×W×L) 52.5 × 84 × 188.5 mm. Traceable certificate of calibration included per OEM. 1-year and 2-year specifications published. |
| Weight | 515 g |