Optris

Optris CT LT Infrared Temperature Sensor

SKU · 44969 · CT LT · Brand: Optris
Available
Quote on request
No online payment · add lines to RFQ BOM

Optris CT LT: −50…975 °C; 8–14 µm; 2:1/15:1/22:1; 150 ms; IP65. RFQ.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Temperature Range
−50…975 °C (LT22); −50…600 °C (LT15 / LT02)
Spectral Range
8–14 μm
Optical Resolution
22:1 (LT22); 15:1 (LT15); 2:1 (LT02)
System Accuracy
±1% or ±1 °C (whichever greater; Tamb 23±5 °C; Tobj>0 °C)
Repeatability
±0.5% or ±0.5 °C
Response Time
150 ms (95%)
Download datasheet PDF

Fits general-purpose non-contact temperature where you need an Optris CT LT—two-piece IR sensor with LT22/LT15/LT02 optics (22:1 / 15:1 / 2:1), spans to about −50…975 °C, ±1% or ±1 °C accuracy class, 150 ms response, and head ambient to 180 °C without cooling on higher optics.

Key advantages

  • Ranges — −50…975 °C (LT22); −50…600 °C (LT15/LT02)
  • Optics — 22:1 / 15:1 / 2:1
  • Accuracy — ±1% or ±1 °C class
  • Response — 150 ms (95%)
  • Head ambient — Up to 180 °C without cooling (LT22/LT15)

Typical applications

  • Process IR spots.
  • OEM packaging.
  • Machine temperature monitoring.

Selection guide

  • Optic ratio
  • Output / fieldbus
  • Head ambient need

Common mistakes

  • Spot > target.
  • Wrong optic for distance.
  • Ignoring head ambient.

When to RFQ / what to tell us

  1. Optic/output.
  2. Quantity, ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU44969
BrandOptris
ModelCT LT
Product typeInfrared Temperature Sensor
Temperature range−50…975 °C (LT22); −50…600 °C (LT15 / LT02)
Spectral range8–14 μm
Optical resolution22:1 (LT22); 15:1 (LT15); 2:1 (LT02)
System accuracy±1% or ±1 °C (whichever greater; Tamb 23±5 °C; Tobj>0 °C)
Repeatability±0.5% or ±0.5 °C
Response time150 ms (95%)
Ambient head−20…180 °C (LT22/LT15); −20…130 °C (LT02) without cooling
Outputs0/4–20 mA, 0–5/10 V, thermocouple J/K; optional USB/RS232/RS485/Profibus/Ethernet

Documents

FAQ

Q: Optics? A: 22:1, 15:1, or 2:1.

Q: Response? A: 150 ms.

Q: What should I send for a quote? A: Optic/output, quantity, ship-to.

Links

ParameterValue
Temperature Range−50…975 °C (LT22); −50…600 °C (LT15 / LT02)
Spectral Range8–14 μm
Optical Resolution22:1 (LT22); 15:1 (LT15); 2:1 (LT02)
System Accuracy±1% or ±1 °C (whichever greater; Tamb 23±5 °C; Tobj>0 °C)
Repeatability±0.5% or ±0.5 °C
Response Time150 ms (95%)
Ambient Head−20…180 °C (LT22/LT15); −20…130 °C (LT02) without cooling
Outputs0/4–20 mA, 0–5/10 V, thermocouple J/K; optional USB/RS232/RS485/Profibus/Ethernet
Power Supply8–36 V DC; max 100 mA
ProtectionIP65 (NEMA-4) sensing head
Emissivity0.100–1.100 adjustable
Weight40 g sensing head / 420 g electronics