Michell Instruments

Michell XTC601 Binary Gas Analyzer

SKU · 42637 · XTC601 · Brand: Michell Instruments
Available
Quote on request
No online payment · add lines to RFQ BOM

Michell XTC601: binary gas analyzer; 0–1% to 0–100%; suppressed. RFQ.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Wetted parts
316L SS, borosilicate glass, platinum, O-rings (Viton/Silicone/Ekraz)
Hazardous area
Ex d / ATEX/IECEx/cQPSus variants; SIL2 capable option
Technology
Thermal conductivity (matched thermistors / Wheatstone bridge)
Measurement Range
Selectable 0…1 % up to 0…100 %; suppressed-zero bands e.g. 50…100 % to 98…100 %
Accuracy H2 He
< ±1 % of range or ±0.05 % H₂/He (whichever greater)
Accuracy Other
< ±2 % of range (other gas mixtures/ranges)
Download datasheet PDF

Fits binary gas mixture analysis where you need a Michell Instruments XTC601—binary gas analyzer covering spans from 0–1% to 0–100% with suppressed high-end ranges such as 50–100% through 98–100% class configurations.

Key advantages

  • Measurand — Binary gas mixture composition
  • Spans — 0–1% to 0–100% class
  • Suppressed — 50–100% to 98–100% class
  • Form — Process binary gas analyzer
  • Platform — Michell XTC601

Typical applications

  • Binary mix monitoring.
  • Purity / balance gas.
  • Process gas quality.

Selection guide

  • Gas pair
  • Span
  • Sample system

Common mistakes

  • Wrong binary pair.
  • Moisture interference.
  • Unsuitable sample flow.

When to RFQ / what to tell us

  1. Gas pair/span.
  2. Quantity, ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU42637
BrandMichell Instruments
ModelXTC601
Product typeBinary Gas Analyzer
TypeBinary gas analyzer (XTC601)
Range0–1% to 0–100%; 50–100% to 98–100% class spans
Measurement typegas
PlatformMichell XTC601

Documents

FAQ

Q: What does it measure? A: Binary gas mixture composition.

Q: Spans? A: 0–1% to 0–100%; suppressed high ranges available.

Q: What should I send for a quote? A: Gas pair/span, quantity, ship-to.

Links

ParameterValue
Wetted parts316L SS, borosilicate glass, platinum, O-rings (Viton/Silicone/Ekraz)
Hazardous areaEx d / ATEX/IECEx/cQPSus variants; SIL2 capable option
TechnologyThermal conductivity (matched thermistors / Wheatstone bridge)
Measurement RangeSelectable 0…1 % up to 0…100 %; suppressed-zero bands e.g. 50…100 % to 98…100 %
Accuracy H2 He< ±1 % of range or ±0.05 % H₂/He (whichever greater)
Accuracy Other< ±2 % of range (other gas mixtures/ranges)
Repeatability±0.2 % of range
Sample Flow Gp100…500 ml/min (0.25…1.0 scfh)
Sample Flow Ex270…330 ml/min Ex version
Sample PressureFixed 0.75…1.5 barA (calibrate at operating pressure); max safe 1 barg
Sample GasNon-condensing; particles <3 μm
Outputs2× 4–20 mA; Modbus RTU RS485; 2 concentration alarm relays
Ingress ProtectionIP66 / NEMA 4 (XTC601); IP55 (XTC501)
Dimensions Xtc601234 × 234 × 172 mm; weight 9.7 kg
Power24 V DC, 1.0 A max