Alpha Omega Instruments

Alpha Omega Instruments ZRO2000 Fast Response Oxygen Analyzer

SKU · 45779 · ZRO2000 · Brand: Alpha Omega Instruments
Available
Quote on request
No online payment · add lines to RFQ BOM

AOI ZRO2000: ZrO2 O₂ analyzer; PPB to 100%; <5 s; dual 4–20 mA; 90–264 VAC. RFQ.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Measuring range
0–100% oxygen (PPB to 100%)
Accuracy
Percent: ±2% of reading or 0.05% absolute (whichever greater); Trace: ±2% of reading or ±0.5 ppm absolute
Output
Two user-scalable 0–20 mA or 4–20 mA; two SPDT Form C alarm relays (10 A @ 30 VDC / 115/230 VAC)
Supply
Universal 90–264 VAC, 50/60 Hz
Display
0.7 in (17.8 mm) 6-digit LCD; 0.01% / 0.001 ppm resolution
Sensor type
Non-depleting zirconium oxide solid-state sensor
Download datasheet PDF

Fits fast oxygen analysis where you need an Alpha Omega Instruments ZRO2000. Non-depleting zirconium oxide analyzer spanning PPB to 100% O₂ with response under 5 seconds at 0.15 LPM over one decade, dual user-scalable 0–20/4–20 mA outputs, and two SPDT Form C alarm relays.

Key advantages

  • Range — PPB to 100% oxygen
  • Sensor — Non-depleting ZrO₂ solid-state
  • Response — <5 s at 0.15 LPM over 1 decade
  • Outputs — Dual scalable 0–20/4–20 mA; 2× SPDT relays
  • Power — Universal 90–264 VAC

Typical applications

  • Fast O₂ process analysis.
  • Inerting / combustion checks.
  • Trace-to-percent O₂ panels.

Selection guide

  • Percent vs trace scaling
  • Alarm setpoints
  • Sample flow ~0.15 LPM class

Common mistakes

  • Wrong sample flow.
  • Liquid in sensor.
  • Ignoring ±2% reading / absolute floor.

When to RFQ / what to tell us

  1. Range/alarms.
  2. Quantity, ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU45779
BrandAlpha Omega Instruments
ModelZRO2000
Product typeOxygen Analyzer
Measuring range0–100% oxygen (PPB to 100%)
AccuracyPercent: ±2% of reading or 0.05% absolute (whichever greater); Trace: ±2% of reading or ±0.5 ppm absolute
Response time<5 seconds at 0.15 LPM over 1 decade
Sensor typeNon-depleting zirconium oxide solid-state sensor
Output signalTwo user-scalable 0–20 mA or 4–20 mA; two SPDT Form C alarm relays (10 A @ 30 VDC / 115/230 VAC)
Supply voltageUniversal 90–264 VAC, 50/60 Hz

Documents

FAQ

Q: Sensor? A: Non-depleting zirconium oxide.

Q: Response? A: <5 s at 0.15 LPM over 1 decade.

Q: What should I send for a quote? A: Range/alarms, quantity, ship-to.

Links

ParameterValue
Measuring range0–100% oxygen (PPB to 100%)
AccuracyPercent: ±2% of reading or 0.05% absolute (whichever greater); Trace: ±2% of reading or ±0.5 ppm absolute
OutputTwo user-scalable 0–20 mA or 4–20 mA; two SPDT Form C alarm relays (10 A @ 30 VDC / 115/230 VAC)
SupplyUniversal 90–264 VAC, 50/60 Hz
Display0.7 in (17.8 mm) 6-digit LCD; 0.01% / 0.001 ppm resolution
Sensor typeNon-depleting zirconium oxide solid-state sensor
Response Time<5 seconds at 0.15 LPM over 1 decade
Operating Temperature50° to 104°F (10° to 40°C)
EnclosureAluminum NEMA 1; 13 × 10.71 × 6.3 in (330 × 272 × 160 mm)
Warranty2 years electronics / 1 year sensor