Panametrics

Panametrics MG101 Portable Hygrometer Calibrator

SKU · 19517 · MG101 · Brand: Panametrics
Available
Quote on request
No online payment · add lines to RFQ BOM

Portable dew-point generator; −75…+10 °Cdp; no power. RFQ + selection help.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Measuring range
−75 °C to +10 °C dew/frost point (−103 °F to 50 °F) at typical ambient
Accuracy
±1 °C (±1.8 °F) dew/frost point
Process connection
1/4 in Swagelok tube fittings
Generation Method
Single- or two-stage gas dilution of saturated + dry streams
Inlet Gas
Nitrogen preferred; must be ≥25 °C drier than driest setpoint
Supply Pressure
55–60 psig (4–5 bar)
Download datasheet PDF

Fits field dew-point verification where you need a Panametrics MG101 portable generator using gas dilution to produce −75 to +10 °C dew/frost point without electrical power.

Key advantages

  • Range — −75 °C to +10 °C dew/frost point at typical ambient
  • Accuracy — ±1 °C (±1.8 °F) dew/frost point
  • Method — Single- or two-stage gas dilution of saturated + dry streams
  • Utilities — Nitrogen preferred; 55–60 psig; no electrical power
  • Flow — Max 19.5 SCFH (9.2 L/min); outlet ambient to 10 psig adjustable

Typical applications

  • Portable verification of dew-point transmitters/hygrometers.
  • Sites with nitrogen available but no power at the test point.
  • Calibration carts needing −75…+10 °Cdp generation.

Selection guide

  • Setpoints — Within −75…+10 °Cdp and inlet dryness (≥25 °C drier than driest setpoint)
  • Gas — Nitrogen preferred at 55–60 psig
  • Connections — 1/4 in Swagelok tube fittings
  • Transport — 25 lb (11.3 kg) dry; 12×18×6 in

Common mistakes

  • Using wet plant air as inlet when dryness requirement is not met.
  • Expecting climatic temperature control — MG101 generates dew/frost point by dilution.
  • Ignoring outlet pressure limit (to 10 psig).

When to RFQ / what to tell us

  1. Target dew-point setpoints; inlet gas available.
  2. Number of DUT connections.
  3. Quantity and ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU19517
BrandPanametrics
ModelPanametrics MG101 Portable Hygrometer Calibration System
Product typeHumidity Calibrator
ModelMG101
Measuring range−75 °C to +10 °C dew/frost point (−103 °F to 50 °F) at typical ambient
Accuracy±1 °C (±1.8 °F) dew/frost point
Generation methodSingle- or two-stage gas dilution of saturated + dry streams
Inlet gasNitrogen preferred; must be ≥25 °C drier than driest setpoint
Supply pressure55–60 psig (4–5 bar)
Max flow rate19.5 SCFH (9.2 L/min)
Outlet pressureAmbient to 10 psig (1 bar), adjustable
PowerNone
Dimensions12 × 18 × 6 in (305 × 457 × 152 mm) W×H×D
Weight25 lb (11.3 kg) dry
Process connection1/4 in Swagelok tube fittings

Documents

FAQ

Q: What range? A: −75 °C to +10 °C dew/frost point at typical ambient.

Q: Does it need power? A: No electrical power required.

Q: What should I send for a quote? A: Setpoints, inlet gas, quantity, ship-to.

Links

ParameterValue
Measuring range−75 °C to +10 °C dew/frost point (−103 °F to 50 °F) at typical ambient
Accuracy±1 °C (±1.8 °F) dew/frost point
Process connection1/4 in Swagelok tube fittings
Generation MethodSingle- or two-stage gas dilution of saturated + dry streams
Inlet GasNitrogen preferred; must be ≥25 °C drier than driest setpoint
Supply Pressure55–60 psig (4–5 bar)
Max Flow Rate19.5 SCFH (9.2 L/min)
Outlet PressureAmbient to 10 psig (1 bar), adjustable
Power RequirementsNone
Dimensions12 × 18 × 6 in (305 × 457 × 152 mm) W×H×D
Weight25 lb (11.3 kg) dry
ApplicationsField calibration/verification of Al-oxide moisture probes and humidity sensors
Special FeaturesInherently no-electric / portable primary moisture generator