Michell Instruments

Michell Instruments Opti-Cal RH & Temperature Calibrator

SKU · 46125 · Opti-Cal RH & Temperature Calibrator · Brand: Michell Instruments
Available
Quote on request
No online payment · add lines to RFQ BOM

Michell Opti-Cal: 10–90% RH; ≤±1% RH (10–70%); +10…+50 °C; ±0.1 °C. RFQ.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Humidity Generated Range
10–90% RH
Humidity Control Accuracy
≤±1% RH (10–70% RH); ≤±1.5% RH (70–90% RH)
Humidity Stability
±0.2% RH (20–80% RH)
Temperature Generated Range
+10 to +50°C (+50 to +122°F); lowest set-point 10°C below ambient
Temperature Accuracy Stability
±0.1°C accuracy; ±0.1°C stability
Chamber Ramp Rate
+20→+40°C: 1.5°C/min; +40→+20°C: 0.7°C/min
Download datasheet PDF

Fits RH probe calibration where you need a Michell Instruments Opti-Cal. Portable humidity and temperature calibrator generating 10–90% RH with control ≤±1% RH (10–70% RH) / ≤±1.5% RH (70–90% RH), ±0.2% RH stability (20–80%), and chamber temperature +10 to +50 °C at ±0.1 °C accuracy/stability.

Key advantages

  • RH generated — 10–90% RH
  • RH control — ≤±1% (10–70%); ≤±1.5% (70–90%)
  • RH stability — ±0.2% RH (20–80%)
  • Temperature — +10 to +50 °C; ±0.1 °C
  • Portable — Chamber calibrator form

Typical applications

  • RH probe calibration.
  • Transmitter checks.
  • Lab / field humidity cal.

Selection guide

  • Probe adapters
  • Setpoint vs ambient
  • Reference probe

Common mistakes

  • Lowest T below ambient−10 °C.
  • Overloaded chamber.
  • Ignoring RH stability band.

When to RFQ / what to tell us

  1. Adapters/options.
  2. Quantity, ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU46125
BrandMichell Instruments
ModelOpti-Cal RH & Temperature Calibrator
Product typeHumidity Calibrator
Humidity generated range10–90% RH
Humidity control accuracy≤±1% RH (10–70% RH); ≤±1.5% RH (70–90% RH)
Humidity stability±0.2% RH (20–80% RH)
Temperature generated range+10 to +50°C (+50 to +122°F); lowest set-point 10°C below ambient
Temperature accuracy/stability±0.1°C accuracy; ±0.1°C stability
Chamber ramp rate+20→+40°C: 1.5°C/min; +40→+20°C: 0.7°C/min

Documents

FAQ

Q: RH range? A: 10–90% RH.

Q: T accuracy? A: ±0.1 °C.

Q: What should I send for a quote? A: Adapters/options, quantity, ship-to.

Links

ParameterValue
Humidity Generated Range10–90% RH
Humidity Control Accuracy≤±1% RH (10–70% RH); ≤±1.5% RH (70–90% RH)
Humidity Stability±0.2% RH (20–80% RH)
Temperature Generated Range+10 to +50°C (+50 to +122°F); lowest set-point 10°C below ambient
Temperature Accuracy Stability±0.1°C accuracy; ±0.1°C stability
Chamber Ramp Rate+20→+40°C: 1.5°C/min; +40→+20°C: 0.7°C/min
Reference AccuracyDew point ±0.2°C; temperature ±0.1°C (integrated Optidew)
Probe PortsUp to 5 ports; sensor body diameters 5–25 mm via adapters
Chamber Volume Dimensions2000 cm³; chamber 105×105×160 mm (W×H×D)
Instrument Dimensions520×290×420 mm (W×H×D)
OutputsAnalog 4–20 mA or 0–20 mA; RS232 @ 9600 baud; volt-free alarm 30 V / 100 mA
Power Weight85–264 V AC, 47/63 Hz, 150 VA; weight 20 kg (44 lb)
ControlManual front-panel or automated time-based set-point programs via PC software
Set Point Resolution0.1 for humidity and temperature