Isotech

Isotech Isorac Model 844 Rack-Mount Thermocouple Reference

SKU · 45115 · Isorac Model 844 Thermocouple Reference Unit · Brand: Isotech
Available
Quote on request
No online payment · add lines to RFQ BOM

Isotech Isorac 844: rack 0 °C reference; to 100 ch; ±0.1 °C/15° ambient. RFQ.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Accuracy
±0.1 °C per 15 °C ambient span
Approvals
Approved for power station use (OEM datasheet)
Reference Temperature
Effectively 0 °C via ambient-following reservoir with compensation signal
Temperature Gradient
±0.1 °C between junctions
Stability
±0.05 °C per 1000 hours
Ambient Range
−20 °C to +70 °C
Download datasheet PDF

Fits rack-mounted thermocouple referencing where you need an Isotech Isorac Model 844—ambient-following 0 °C compensation reference with up to 100 double-junction channels, ±0.1 °C accuracy per 15 °C ambient span, and 19-inch rack dimensions.

Key advantages

  • Reference — Effectively 0 °C via ambient-following reservoir
  • Channels — 0 to 100 double-junction
  • Accuracy — ±0.1 °C per 15 °C ambient span
  • Stability — ±0.05 °C per 1000 hours
  • Form — Rack-mount Isorac

Typical applications

  • Rack CJC systems.
  • Multi-channel DAQ references.
  • Control-room TC panels.

Selection guide

  • Channel count / depth
  • Compensation output type
  • AC vs low-level DC power

Common mistakes

  • Wrong rack depth for channel count.
  • Skipping 10 min stabilisation.
  • Ambient outside −20…+70 °C.

When to RFQ / what to tell us

  1. Channels/output/power.
  2. Quantity, ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU45115
BrandIsotech
ModelIsorac Model 844 Thermocouple Reference Unit
Product typeThermocouple Reference Unit
Reference temperatureEffectively 0 °C via ambient-following reservoir with compensation signal
Temperature gradient±0.1 °C between junctions
Stability±0.05 °C per 1000 hours
Accuracy±0.1 °C per 15 °C ambient span
Thermocouple capacity0 to 100 channels double-junction referencing
Compensation outputDC mV, 4–20 mA, or platinum resistance thermometer (order-specific)
Power10 W typical; 100–130 or 208–240 VAC 50/60 Hz; low-level DC 6 VA option
DimensionsUp to 50 ch: H 255 × W 483 × D 312 mm (17.2 kg); 50–100 ch: H 309 × W 483 × D 312 mm (20.4 kg)

Documents

FAQ

Q: Form? A: Rack-mount Isorac.

Q: Channels? A: Up to 100.

Q: What should I send for a quote? A: Channels/output/power, quantity, ship-to.

Links

ParameterValue
Accuracy±0.1 °C per 15 °C ambient span
ApprovalsApproved for power station use (OEM datasheet)
Reference TemperatureEffectively 0 °C via ambient-following reservoir with compensation signal
Temperature Gradient±0.1 °C between junctions
Stability±0.05 °C per 1000 hours
Ambient Range−20 °C to +70 °C
Stabilisation Time10 minutes
Thermocouple Capacity0 to 100 channels double-junction referencing
Compensation OutputDC mV, 4–20 mA, or platinum resistance thermometer (order-specific)
Power10 W typical; 100–130 or 208–240 VAC 50/60 Hz; low-level DC 6 VA option
DimensionsUp to 50 ch: H 255 × W 483 × D 312 mm (17.2 kg); 50–100 ch: H 309 × W 483 × D 312 mm (20.4 kg)
Special Features19-inch rack-mount ambient referencing; approved for power-station use