Time Electronics

Time Electronics 1006 DC Millivolt Source

SKU · 25574 · 1006 · Brand: Time Electronics
Available
Quote on request
No online payment · add lines to RFQ BOM

Portable mV source; to 999.9 mV; 1 µV res; ±0.02%. RFQ + selection help.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Accuracy
±(0.02 % of setting + 0.02 % of range + 1 µV)
Type
Portable DC millivolt source — Time Electronics 1006
Output Ranges
0–999.9 mV (0.1 mV); 0–99.99 mV (10 µV); 0–9.999 mV (1 µV)
Max Output Current
20 mA on 1 V and 100 mV ranges
Output Resistance
<0.2 Ω (1 V/100 mV); 1 Ω (10 mV range)
Polarity
Positive / negative / centre-off switch
Download datasheet PDF

Fits portable low-level DC sourcing where you need a Time Electronics 1006 covering 0–999.9 mV with 1 µV resolution on the lowest range and ±0.02% class accuracy.

Key advantages

  • Ranges — 0–999.9 mV / 0–99.99 mV / 0–9.999 mV
  • Resolution — Down to 1 µV
  • Accuracy — ±(0.02% of setting + 0.02% of range + 1 µV)
  • Drive — 20 mA max on 1 V and 100 mV ranges; Rout <0.2 Ω
  • Power — 6× AA; ~100 h typical; short-circuit protected

Typical applications

  • Thermocouple instrument simulation (mV).
  • Low-level transmitter checks.
  • Portable precision mV sourcing.

Selection guide

  • Lowest range — 1 µV need
  • Output current — 20 mA class
  • Polarity — + / − / centre-off
  • Battery life — ~100 h field use

Common mistakes

  • Overloading beyond 20 mA.
  • Ignoring output resistance on 10 mV range (1 Ω).
  • Leaving polarity centre-off during test.

When to RFQ / what to tell us

  1. Quantity; ship-to; accessories.

Request a Quote · Get selection help

Specs

SpecValue
SKU25574
BrandTime Electronics
Model1006
Product typeDC Millivolt Source
TypePortable DC millivolt source — Time Electronics 1006
Output ranges0–999.9 mV (0.1 mV); 0–99.99 mV (10 µV); 0–9.999 mV (1 µV)
Accuracy±(0.02 % of setting + 0.02 % of range + 1 µV)
Max output current20 mA on 1 V and 100 mV ranges
Output resistance<0.2 Ω (1 V/100 mV); 1 Ω (10 mV range)
PolarityPositive / negative / centre-off switch
ProtectionShort-circuit and overload protected; safety terminals
Power6× AA; ~100 h typical battery life; removable protective boot

Documents

FAQ

Q: Best resolution? A: 1 µV on the 0–9.999 mV range.

Q: Accuracy? A: ±(0.02% of setting + 0.02% of range + 1 µV).

Q: What should I send for a quote? A: Quantity and ship-to.

Links

ParameterValue
Accuracy±(0.02 % of setting + 0.02 % of range + 1 µV)
TypePortable DC millivolt source — Time Electronics 1006
Output Ranges0–999.9 mV (0.1 mV); 0–99.99 mV (10 µV); 0–9.999 mV (1 µV)
Max Output Current20 mA on 1 V and 100 mV ranges
Output Resistance<0.2 Ω (1 V/100 mV); 1 Ω (10 mV range)
PolarityPositive / negative / centre-off switch
ProtectionShort-circuit and overload protected; safety terminals
Power6× AA; ~100 h typical battery life; removable protective boot
ApplicationsThermocouple simulation, chart recorder / DMM / A-D calibration
Document1006 datasheet + 1006/1007 user manual