Additel

Additel 209/210 Series Loop Calibrator

SKU · 43569 · 209 and 210 Series · Brand: Additel
Available
Quote on request
No online payment · add lines to RFQ BOM

Handheld loop calibrator. 0–24 mA measure/source/simulate, 24 V loop, HART 250 Ω. ADT209 0.03% / ADT210 0.01%. RFQ + help.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Measuring range
Voltage DC 0–30 V; Current DC 0–24 mA (measure/source/simulate); switch test
Accuracy
ADT209 0.03%RD + 2 counts; ADT210 0.01%RD + 2 counts (10–30 °C, 1 year)
Output
mA source/sink; 24 V loop power; HART 250 Ω resistor in series with 24 V loop
Supply
One 9 V alkaline battery; optional DC 9 V adapter (9812)
Electrical connection
Test leads / alligator clips
Protection
IP54 (per user manual)
Download datasheet PDF

Fits handheld 4–20 mA loop check and simulation when you need measure / source / sink, 24 V loop power, and an in-series HART 250 Ω resistor — choose ADT209 (0.03%RD) or ADT210 (0.01%RD).

Key advantages

  • 0–24 mA full set — Measure / source / simulate plus switch test.
  • ADT209 or ADT210 — 0.03%RD or 0.01%RD (10–30 °C, 1 year).
  • 24 V + HART 250 Ω — Loop power with in-series HART resistor.
  • IP54 handheld — 163×83×41 mm, 350 g, one 9 V alkaline.

Typical applications

  • Commissioning and troubleshooting 4–20 mA transmitter loops.
  • Simulating transmitters into PLC/DCS inputs with % span readout.
  • Quick HART-device loop checks that need a 250 Ω resistor in series.

Selection guide

  • Ranges — Voltage DC 0–30 V; Current DC 0–24 mA (measure / source / simulate); switch test.
  • Accuracy — ADT209 0.03%RD + 2 counts; ADT210 0.01%RD + 2 counts (10–30 °C, 1 year).
  • Loop features — mA source/sink; 24 V loop power; HART 250 Ω resistor in series with 24 V loop.
  • Supply — One 9 V alkaline; optional DC 9 V adapter (9812). Battery life typical: output 18 h / measure 50 h.
  • Display — VA LCD 52×52 mm; simultaneous mA and % span.
  • IP / ambient — IP54; working −10…+50 °C, 95% RH; storage −20…+70 °C.
  • Approvals — CE; ISO 17025 Accredited Calibration with data.

Common mistakes

  • Ordering ADT209 when the procedure needs ADT210’s 0.01%RD — accuracy is model-specific.
  • Expecting HART communicator functions — unit provides a 250 Ω resistor in series, not a full HART master tool.
  • Using outside −10…+50 °C working range or ignoring the 10–30 °C accuracy window.
  • Assuming IP54 covers immersion — handheld splash/dust rating only.
  • Skipping the ISO 17025 calibration-with-data option when QA requires it on receipt.

When to RFQ / what to tell us

  1. Model: ADT209 (0.03%RD) vs ADT210 (0.01%RD).
  2. Primary use (measure/source/simulate), ISO 17025 yes/no, quantity, ship-to.
  3. Any need for a full HART communicator instead of the 250 Ω resistor path.

Request a Quote · Get selection help

Specs

SpecValue
SKU43569
BrandAdditel
Model209 and 210 Series
Product typecalibrator
Measuring rangeVoltage DC 0–30 V; Current DC 0–24 mA (measure/source/simulate); switch test
AccuracyADT209 0.03%RD + 2 counts; ADT210 0.01%RD + 2 counts (10–30 °C, 1 year)
Output signalmA source/sink; 24 V loop power; HART 250 Ω resistor in series with 24 V loop
Supply voltageOne 9 V alkaline battery; optional DC 9 V adapter (9812)
Electrical connectionTest leads / alligator clips
IP ratingIP54 (per user manual)
Ambient temperatureWorking −10…+50 °C, 95% RH; storage −20…+70 °C
DisplayVA LCD 52×52 mm; simultaneous mA and % span
ApprovalsCE (ADT209); CE (ADT210); ISO 17025 Accredited Calibration with data
MountingHandheld
Weight350 g
NotesSize 163×83×41 mm. Over-voltage protection 30 V DC (240 V AC). Battery life typical: output 18 h / measure 50 h.

Documents

FAQ

Q: ADT209 or ADT210 — which accuracy? A: ADT209 is 0.03%RD + 2 counts; ADT210 is 0.01%RD + 2 counts (both 10–30 °C, 1 year). Pick from the tolerance your procedure allows.

Q: Does this replace a HART communicator? A: No. It provides 24 V loop power with a HART 250 Ω resistor in series so a HART device can talk on the loop. Full HART configuration needs a separate communicator/app.

Q: What calibration paperwork is available? A: Datasheet lists ISO 17025 Accredited Calibration with data (plus CE on both models). Ask for it explicitly on the RFQ if required.

Q: What should I send for a quote? A: ADT209 vs ADT210, functions needed, adapter/cal cert yes/no, quantity. Compatible with systems using Additel 209/210 if you need a drop-in check — we do not claim authorized-dealer status.

Links

ParameterValue
Measuring rangeVoltage DC 0–30 V; Current DC 0–24 mA (measure/source/simulate); switch test
AccuracyADT209 0.03%RD + 2 counts; ADT210 0.01%RD + 2 counts (10–30 °C, 1 year)
OutputmA source/sink; 24 V loop power; HART 250 Ω resistor in series with 24 V loop
SupplyOne 9 V alkaline battery; optional DC 9 V adapter (9812)
Electrical connectionTest leads / alligator clips
ProtectionIP54 (per user manual)
Ambient temperatureWorking −10…+50 °C, 95% RH; storage −20…+70 °C
DisplayVA LCD 52×52 mm; simultaneous mA and % span
ApprovalsCE (ADT209); CE (ADT210); ISO 17025 Accredited Calibration with data
MountingHandheld
NotesSize 163×83×41 mm. Over-voltage protection 30 V DC (240 V AC). Battery life typical: output 18 h / measure 50 h.
Weight350 g