Time Electronics

Time Electronics 5045 Oscilloscope Calibrator

SKU · 12347 · 5045 · Brand: Time Electronics
Available
Quote on request
No online payment · add lines to RFQ BOM

Scope/timer/frequency calibrator. To 100 MHz; <300 ps rise. RFQ + selection help.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Amplitude Range
1 mV to 220 V square wave/DC (2 V max into 50 Ω)
Frequency Range
0.1 Hz to 100 MHz
Time Markers
10 s to 10 ns
Fast Rise
< 300 ps (bandwidth testing to ~600 MHz)
Timing Accuracy
0.1 ppm (TCXO); optional rubidium reference
Interfaces
Front panel; PC virtual control; RS-232, USB, GPIB
Download datasheet PDF

Fits metrology labs calibrating scopes and timer/counters where you need a Time Electronics 5045 with amplitude, frequency, period, and fast-rise outputs.

Key advantages

  • Amplitude — 1 mV to 220 V square wave/DC (2 V max into 50 Ω)
  • Frequency — 0.1 Hz to 100 MHz
  • Time markers — 10 s to 10 ns
  • Fast rise — < 300 ps (bandwidth testing class)
  • Interfaces / options — per OEM datasheet interfaces; EasyCal-oriented lab use

Typical applications

  • Oscilloscope amplitude and timebase calibration.
  • Timer-counter / frequency-meter calibration.
  • Bandwidth checks using the fast-rise output.

Selection guide

  • Amplitude / load — high-Z vs 50 Ω limits (2 V max into 50 Ω)
  • Frequency / markers — 0.1 Hz–100 MHz; 10 s–10 ns markers
  • Options / interfaces — lock from OEM datasheet on RFQ

Common mistakes

  • Driving 50 Ω loads above the 2 V max amplitude limit.
  • Assuming every option card is included — confirm interfaces on RFQ.
  • Using as a general multifunction process calibrator — 5045 is scope/timer focused.

When to RFQ / what to tell us

  1. Required amplitude/frequency/marker ranges.
  2. Interface/options; quantity; ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU12347
BrandTime Electronics
Model5045 Oscilloscope Calibrator
Product typeOscilloscope Calibrator
Amplitude range1 mV to 220 V square wave/DC (2 V max into 50 Ω)
Frequency range0.1 Hz to 100 MHz
Time markers10 s to 10 ns
Fast rise< 300 ps (bandwidth testing to ~600 MHz)
Timing accuracy0.1 ppm (TCXO); optional rubidium reference
InterfacesFront panel; PC virtual control; RS-232, USB, GPIB
Options2.2 GHz levelled sine; rubidium reference; current probe adaptor
Weight8.2 kg (18 lb)

Documents

FAQ

Q: Frequency range? A: 0.1 Hz to 100 MHz.

Q: Fast-rise spec? A: < 300 ps class for bandwidth testing.

Q: What should I send for a quote? A: Ranges needed, options, quantity, ship-to.

Links

ParameterValue
Amplitude Range1 mV to 220 V square wave/DC (2 V max into 50 Ω)
Frequency Range0.1 Hz to 100 MHz
Time Markers10 s to 10 ns
Fast Rise< 300 ps (bandwidth testing to ~600 MHz)
Timing Accuracy0.1 ppm (TCXO); optional rubidium reference
InterfacesFront panel; PC virtual control; RS-232, USB, GPIB
Options2.2 GHz levelled sine; rubidium reference; current probe adaptor
Weight8.2 kg (18 lb)
Deviation±9.99% amplitude deviation with direct error readout
SoftwareWindows virtual control; EasyCal optional