RFQ cart holds BOM lines only. Submit for engineering quote.
Key parameters
Hazardous area
ATEX/IECEx Ex ia; CSA Class I Div 1; intrinsically safe
Pressure Ranges
Vacuum to 15,000 psi (model-dependent); gauge and absolute versions
Accuracy Standard
0–20% FS: ±0.02% FS; 20–110% FS: ±0.1% of reading (−10 to 50 °C)
Accuracy Options
−S5 ±0.05% FS; −S2 ±0.02% FS (15–25 °C)
Operating Temperature
−10 to 50 °C; accuracy unaffected across range when zeroed
Protection
IP66 / IP67
Fits field pressure test and calibration where you need an intrinsically safe Crystal XP2i digital gauge from vacuum to 15,000 psi class.
Key advantages
- IS digital test gauge — Intrinsically safe digital pressure gauge (Crystal / AMETEK Calibration)
- Ranges — Vacuum to 15,000 psi (model-dependent); gauge and absolute versions
- Accuracy — 0–20% FS: ±0.02% FS; 20–110% FS: ±0.1% of reading (−10 to 50 °C); −S5 / −S2 options
- Hazardous area — ATEX/IECEx Ex ia; CSA Class I Div 1
- Features — IP66/IP67; PSV mode; peak/valley; optional DataLoggerXP (32,000 pts); ISO 17025 certificate included
Typical applications
- Field pressure calibration and PSV testing in hazardous areas.
- Marine-grade enclosures and corrosion-prone sites.
- Logging with DataLoggerXP option.
Selection guide
- Pressure range — vacuum to 15,000 psi class; gauge vs absolute
- Accuracy options — standard vs −S5 / −S2
- Connection — CPF / NPT by range
- Logging — DataLoggerXP yes/no
Common mistakes
- Ordering a range far above working span — resolution suffers.
- Assuming −S2 accuracy without the option and 15–25 °C window.
- Using non-approved AA cells in ATEX/IECEx/CSA service.
When to RFQ / what to tell us
- Pressure range and gauge vs absolute.
- Accuracy option; logging; connection.
- Quantity; ship-to.
Request a Quote · Get selection help
Specs
| Spec | Value |
|---|---|
| SKU | 12351 |
| Brand | Crystal Engineering |
| Model | XP2i Digital Pressure Gauge |
| Product type | Digital Pressure Gauge |
| Pressure ranges | Vacuum to 15,000 psi (model-dependent); gauge and absolute versions |
| Accuracy (standard) | 0–20% FS: ±0.02% FS; 20–110% FS: ±0.1% of reading (−10 to 50 °C) |
| Accuracy options | −S5 ±0.05% FS; −S2 ±0.02% FS (15–25 °C) |
| Operating temperature | −10 to 50 °C; accuracy unaffected across range when zeroed |
| Protection | IP66 / IP67 |
| Hazardous area | ATEX/IECEx Ex ia; CSA Class I Div 1; intrinsically safe |
| Power | 3× AA (ATEX/IECEx/CSA approved types); long battery life |
| Features | PSV mode; peak/valley; optional DataLoggerXP (32,000 pts); marine-grade enclosure |
| Sensor / connection | All-welded 316 SS sensor; CPF / NPT process connections by range |
Documents
FAQ
Q: What ranges? A: Vacuum to 15,000 psi, model-dependent; gauge and absolute versions.
Q: IS approvals? A: ATEX/IECEx Ex ia; CSA Class I Div 1.
Q: What should I send for a quote? A: Range, accuracy option, logging, connection, quantity, ship-to.
Links
| Parameter | Value |
|---|---|
| Hazardous area | ATEX/IECEx Ex ia; CSA Class I Div 1; intrinsically safe |
| Pressure Ranges | Vacuum to 15,000 psi (model-dependent); gauge and absolute versions |
| Accuracy Standard | 0–20% FS: ±0.02% FS; 20–110% FS: ±0.1% of reading (−10 to 50 °C) |
| Accuracy Options | −S5 ±0.05% FS; −S2 ±0.02% FS (15–25 °C) |
| Operating Temperature | −10 to 50 °C; accuracy unaffected across range when zeroed |
| Protection | IP66 / IP67 |
| Power | 3× AA (ATEX/IECEx/CSA approved types); long battery life |
| Features | PSV mode; peak/valley; optional DataLoggerXP (32,000 pts); marine-grade enclosure |
| Sensor | All-welded 316 SS sensor; CPF / NPT process connections by range |
| Calibration | ISO 17025 calibration certificate included |