DH-Budenberg

DH-Budenberg CPG2500 Precision Pressure Indicator

SKU · 12352 · CPG2500 · Brand: DH-Budenberg
Available
Quote on request
No online payment · add lines to RFQ BOM

Lab pressure indicator. To 0.008% IntelliScale; up to 3 channels. RFQ + selection help.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Accuracy
Down to 0.008 % of IntelliScale (IS); precision 0.004 % FS
Display
7 in colour LCD touchscreen; resolution selectable 4–7 digits
Type
Precision digital pressure indicator / calibrator reference (CPG2500; DH-Budenberg/Mensor/WIKA line)
Pressure Ranges
0.36…42,000 psi (25 mbar…2,890 bar) depending on transducers
Channels
Up to 3 transducer channels (2 internal + 1 external); removable/interchangeable
Communications
RS-232, IEEE-488, USB, Ethernet
Download datasheet PDF

Fits laboratory pressure reference work where you need a CPG2500 precision digital indicator with IntelliScale accuracy and interchangeable transducers (DH-Budenberg / Mensor / WIKA line).

Key advantages

  • Precision indicator — Precision digital pressure indicator / calibrator reference (CPG2500)
  • Ranges — 0.36…42,000 psi (25 mbar…2,890 bar) depending on transducers
  • Accuracy — Down to 0.008 % of IntelliScale (IS); precision 0.004 % FS
  • Channels — Up to 3 transducer channels (2 internal + 1 external); removable/interchangeable
  • I/O — 7 in colour LCD touchscreen; RS-232, IEEE-488, USB, Ethernet

Typical applications

  • Lab pressure reference and indicator calibrations.
  • Multi-channel setups with internal plus external CPT transducers.
  • Automated benches needing IEEE-488 / Ethernet remote control.

Selection guide

  • Transducer ranges — internal/external CPT9000 / CPT6100 / CPT6180 class
  • Accuracy / IntelliScale — lock required IS class
  • Channel count — 1–3 channels
  • Communications — RS-232 / GPIB / USB / Ethernet

Common mistakes

  • Ordering without specifying transducer ranges — CPG2500 is a platform.
  • Ignoring ~15 minute warmup in uncertainty budgets.
  • Assuming every external sensor is included — CPT sensors are separate.

When to RFQ / what to tell us

  1. Transducer ranges and channel count.
  2. Communications; quantity; ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU12352
BrandDH-Budenberg
ModelCPG2500
Product typePrecision Pressure Indicator
TypePrecision digital pressure indicator / calibrator reference (CPG2500; DH-Budenberg/Mensor/WIKA line)
Pressure ranges0.36…42,000 psi (25 mbar…2,890 bar) depending on transducers
AccuracyDown to 0.008 % of IntelliScale (IS); precision 0.004 % FS
ChannelsUp to 3 transducer channels (2 internal + 1 external); removable/interchangeable
Display7 in colour LCD touchscreen; resolution selectable 4–7 digits
CommunicationsRS-232, IEEE-488, USB, Ethernet
External sensorsMensor CPT9000 / CPT6100 / CPT6180 compatible
WarmupApproximately 15 minutes

Documents

FAQ

Q: What accuracy? A: Down to 0.008% of IntelliScale; precision 0.004% FS class.

Q: How many channels? A: Up to 3 (2 internal + 1 external), interchangeable.

Q: What should I send for a quote? A: Ranges/channels, communications, quantity, ship-to.

Links

ParameterValue
AccuracyDown to 0.008 % of IntelliScale (IS); precision 0.004 % FS
Display7 in colour LCD touchscreen; resolution selectable 4–7 digits
TypePrecision digital pressure indicator / calibrator reference (CPG2500; DH-Budenberg/Mensor/WIKA line)
Pressure Ranges0.36…42,000 psi (25 mbar…2,890 bar) depending on transducers
ChannelsUp to 3 transducer channels (2 internal + 1 external); removable/interchangeable
CommunicationsRS-232, IEEE-488, USB, Ethernet
External SensorsMensor CPT9000 / CPT6100 / CPT6180 compatible
WarmupApproximately 15 minutes
WeightApprox. 12.5 lb / 5.7 kg with internal options
DocumentWIKA/Mensor data sheet CT 25.02 + CPG2500 operating instructions