Image unavailable
Image unavailable
Additel
Additel ADT 793 Pressure Controller
Quote on request
No online payment · add lines to RFQ BOM
Additel ADT793: to 15,000 psi; 0.01% RD/FS or 0.02% FS; precision to 0.004% FS. RFQ.
Download datasheet PDF
RFQ cart holds BOM lines only. Submit for engineering quote.
Key parameters
Measuring range
0 to 15,000 psi (1,035 bar); ADT151 modules from −14.5~3,000 psi to 0~15,000 psi
Accuracy
0.01% RD or 0.01% FS or 0.02% FS (module selectable); precision 0.008% RD / 0.007% FS / 0.004% FS
Output
Automated hydraulic pressure control; gauge/absolute (absolute needs ADT151-BP)
Supply
Instrument AC power per datasheet; built-in hydraulic pump (no external pressure source)
Process connection
1/4 BSP M
Pressure type
gauge, absolute
Fits high-pressure automated control where you need an Additel ADT 793. Controller to 15,000 psi (1,035 bar) with ADT151 modules from about −14.5~3,000 psi to 0~15,000 psi, accuracy 0.01% RD or 0.01% FS or 0.02% FS selectable, and precision classes down to 0.008% RD / 0.007% FS / 0.004% FS.
Key advantages
- Range — To 15,000 psi (1,035 bar)
- Accuracy — 0.01% RD / 0.01% FS / 0.02% FS
- Precision — To 0.004% FS class
- Modules — ADT151 family
- Use — High-pressure control
Typical applications
- High-pressure cal benches.
- Hydraulic automated control.
- Precision module systems.
Selection guide
- Module set
- Accuracy/precision class
- Media
Common mistakes
- Module below needed URL.
- Media contamination.
- Confusing RD vs FS.
When to RFQ / what to tell us
- Modules/options.
- Quantity, ship-to.
Request a Quote · Get selection help
Specs
| Spec | Value |
|---|---|
| SKU | 69115 |
| Brand | Additel |
| Model | ADT 793 |
| Product type | Pressure Controller |
| Measuring range | 0 to 15,000 psi (1,035 bar); ADT151 modules from −14.5~3,000 psi to 0~15,000 psi |
| Accuracy | 0.01% RD or 0.01% FS or 0.02% FS (module selectable); precision 0.008% RD / 0.007% FS / 0.004% FS |
| Range min | 0 |
| Range max | 15000 |
| Range unit | psi |
Documents
FAQ
Q: Max? A: 15,000 psi.
Q: Accuracy? A: 0.01% RD / 0.01% FS / 0.02% FS.
Q: What should I send for a quote? A: Modules/options, quantity, ship-to.
Links
| Parameter | Value |
|---|---|
| Measuring range | 0 to 15,000 psi (1,035 bar); ADT151 modules from −14.5~3,000 psi to 0~15,000 psi |
| Accuracy | 0.01% RD or 0.01% FS or 0.02% FS (module selectable); precision 0.008% RD / 0.007% FS / 0.004% FS |
| Output | Automated hydraulic pressure control; gauge/absolute (absolute needs ADT151-BP) |
| Supply | Instrument AC power per datasheet; built-in hydraulic pump (no external pressure source) |
| Process connection | 1/4 BSP M |
| Pressure type | gauge, absolute |
| Display | Touchscreen controller (family shared with 773/783) |
| Communication | LAN, USB, RS232, Ethernet/Wi-Fi; GPIB/SCPI/Modbus per manuals |
| Approvals | CE and regional markings per datasheet |
| Control Stability | <0.005% FS, typically 0.003% FS |
| Control Response Time | <30 seconds (typical 20% step) |
| Media | Diethylhexyl Sebacate (ADT793); deionized/distilled/pure water (ADT793W) |
| Reservoir Volume | Built-in 800 mL; external reservoir optional |
| Maximum Load Volume | <10,000 psi: max 80 mL (50 mL recommend); 10,000–15,000 psi: max 60 mL (35 mL recommend) |