Image unavailable
Image unavailable
Ametek Crystal
Ametek Crystal HPC40 Pressure Calibrator
Quote on request
No online payment · add lines to RFQ BOM
Crystal HPC40: to 15k psi; ±0.01% FS / ±0.035% rdg; mA/V/24V. RFQ.
Download datasheet PDF
RFQ cart holds BOM lines only. Submit for engineering quote.
Key parameters
Accuracy Gauge 18 28C
0–30% FS ±0.01% FS; 30–110% of range ±0.035% of reading
Accuracy Gauge M20 50C
0–30% FS ±0.015% FS; 30–110% of range ±0.050% of reading (full temp compensation)
Pressure Range
Built-in sensors up to 15,000 psi / 1000 bar; gauge/absolute/differential
Electrical
Measure/source mA; VDC in; switch test; internal 24 VDC loop supply
Temperature
RTD input with True Ohm technology
Modules
Optional APM CPF Advanced Pressure Modules (±0.025% rdg class)
Fits portable pressure calibration where you need an Ametek Crystal HPC40—built-in sensors to 15,000 psi / 1000 bar with gauge accuracy 0–30% FS ±0.01% FS and 30–110% of range ±0.035% of reading (18–28 °C), plus mA measure/source, VDC in, switch test, and internal 24 VDC loop supply.
Key advantages
- Pressure — Built-in to 15,000 psi / 1000 bar
- Accuracy (18–28 °C) — 0–30% FS ±0.01% FS; 30–110% ±0.035% rdg
- Temp compensated — Wider band specs to −20…50 °C
- Electrical — mA measure/source; VDC in; switch test; 24 V loop
- Modes — Gauge / absolute / differential class
Typical applications
- Transmitter pressure cal.
- Switch testing.
- Field pressure loops.
Selection guide
- Built-in range
- Pump / hose kit
- Intrinsic safety need (other models)
Common mistakes
- Overranging.
- Wrong media.
- Ignoring temperature band for accuracy.
When to RFQ / what to tell us
- Range, accessories.
- Quantity, ship-to.
Request a Quote · Get selection help
Specs
| Spec | Value |
|---|---|
| SKU | 42145 |
| Brand | Ametek Crystal |
| Model | Crystal HPC40 Pressure Calibrator |
| Product type | Pressure Calibrator |
| Accuracy gauge 18-28C | 0–30% FS ±0.01% FS; 30–110% of range ±0.035% of reading |
| Accuracy gauge -20-50C | 0–30% FS ±0.015% FS; 30–110% of range ±0.050% of reading (full temp compensation) |
| Pressure range | Built-in sensors up to 15,000 psi / 1000 bar; gauge/absolute/differential |
| Electrical | Measure/source mA; VDC in; switch test; internal 24 VDC loop supply |
Documents
FAQ
Q: Max pressure? A: Built-in sensors up to 15,000 psi / 1000 bar.
Q: Electrical? A: mA measure/source, VDC in, switch test, 24 V loop.
Q: What should I send for a quote? A: Range/accessories, quantity, ship-to.
Links
| Parameter | Value |
|---|---|
| Accuracy Gauge 18 28C | 0–30% FS ±0.01% FS; 30–110% of range ±0.035% of reading |
| Accuracy Gauge M20 50C | 0–30% FS ±0.015% FS; 30–110% of range ±0.050% of reading (full temp compensation) |
| Pressure Range | Built-in sensors up to 15,000 psi / 1000 bar; gauge/absolute/differential |
| Electrical | Measure/source mA; VDC in; switch test; internal 24 VDC loop supply |
| Temperature | RTD input with True Ohm technology |
| Modules | Optional APM CPF Advanced Pressure Modules (±0.025% rdg class) |
| Baro Option | Toggle gauge/absolute; BARO accuracy ±0.00725 psi / ±0.5 mbar |
| Connection | Crystal CPF female; 316 SST wetted; silicone oil fill |
| Power | 4×AA or rechargeable options; mini-USB |
| Models | HPC41 (single) / HPC42 (dual sensor) |
| Delivery | ISO 17025 NIST-traceable certificate, adapters, boot, leads |
| Application | Handheld pressure + mA loop calibrator for field/custody transfer |