Image unavailable
Image unavailable
Ametek Crystal
Ametek Crystal 30 Series Digital Pressure Calibrator
Quote on request
No online payment · add lines to RFQ BOM
Crystal 30 Series: IS digital P cal; to 5000 psi; mA ±0.025%+floor. RFQ model.
Download datasheet PDF
RFQ cart holds BOM lines only. Submit for engineering quote.
Key parameters
Display
2-line 8-character LCD; ~2 readings/s (adjustable)
Type
Ultra-compact intrinsically safe digital pressure calibrator
Accuracy Pressure
±(0.05% of reading + floor term); floor 0.005% FS or display resolution
Pressure Range
Vacuum and pressure up to 5000 psi (300 bar) depending on model
Accuracy Ma
±(0.025% of reading + 0.001 mA); range 0–55 mA
Ports
Dual 1/8 in FNPT pressure ports; 4 mm banana jacks for mA
Fits compact intrinsically safe pressure calibration where you need an Ametek Crystal 30 Series—ultra-compact digital pressure calibrator with accuracy ±(0.05% of reading + floor), vacuum and pressure up to 5000 psi (300 bar) by model, and mA accuracy ±(0.025% of reading + 0.001 mA) over 0–55 mA.
Key advantages
- Pressure accuracy — ±(0.05% of reading + floor term)
- Range — Vacuum and pressure up to 5000 psi (300 bar) by model
- mA — ±(0.025% of reading + 0.001 mA); 0–55 mA
- Form — Ultra-compact intrinsically safe calibrator
- Platform — Crystal 30 Series
Typical applications
- Hazloc pressure transmitter cal.
- Field IS gauge checks.
- Compact loop mA + pressure.
Selection guide
- Pressure range model
- IS documentation
- Pump / hose kit
Common mistakes
- Overranging sensor.
- Wrong media.
- Ignoring IS battery rules.
When to RFQ / what to tell us
- Model/range.
- Quantity, ship-to.
Request a Quote · Get selection help
Specs
| Spec | Value |
|---|---|
| SKU | 42146 |
| Brand | Ametek Crystal |
| Model | Crystal 30 Series Digital Pressure Calibrator |
| Product type | Pressure Calibrator |
| Type | Ultra-compact intrinsically safe digital pressure calibrator |
| Accuracy pressure | ±(0.05% of reading + floor term); floor 0.005% FS or display resolution |
| Pressure range | Vacuum and pressure up to 5000 psi (300 bar) depending on model |
| Accuracy mA | ±(0.025% of reading + 0.001 mA); range 0–55 mA |
Documents
FAQ
Q: Max pressure? A: Up to 5000 psi (300 bar) depending on model.
Q: mA accuracy? A: ±(0.025% of reading + 0.001 mA).
Q: What should I send for a quote? A: Model/range, quantity, ship-to.
Links
| Parameter | Value |
|---|---|
| Display | 2-line 8-character LCD; ~2 readings/s (adjustable) |
| Type | Ultra-compact intrinsically safe digital pressure calibrator |
| Accuracy Pressure | ±(0.05% of reading + floor term); floor 0.005% FS or display resolution |
| Pressure Range | Vacuum and pressure up to 5000 psi (300 bar) depending on model |
| Accuracy Ma | ±(0.025% of reading + 0.001 mA); range 0–55 mA |
| Ports | Dual 1/8 in FNPT pressure ports; 4 mm banana jacks for mA |
| Power | Single 9 V battery; ~90 hours typical; micro-USB interface |
| Operating Temperature | 0…50 °C (32…122 °F); no accuracy shift over range when zeroed |
| Certifications | ATEX/IECEx Ex ia IIC T4; CSA Class I Div 1; DNV marine; EMC/PED |
| Size | Approx. 137×70×29 mm; pocket-sized |
| Application | Field transmitter calibration; pressure + mA indication |
| Document | 30 Series psi datasheet + 30 Series operation manual |