Ashcroft

Ashcroft ATE-2 Handheld Calibrator

SKU · 42206 · ATE-2 · Brand: Ashcroft
Available
Quote on request
No online payment · add lines to RFQ BOM

Ashcroft ATE-2: dual AM2 modules; 0.25 inH2O–10k psi; mA/V; IP65. RFQ.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Measuring range
Pressure modules 0.25 in H2O to 10,000 psi; electrical 0–20 mA or 0–30 Vdc
Accuracy
Pressure modules ±0.025% to ±0.1% of span; electrical 0–10 Vdc ±0.025% span, 10–30 Vdc ±0.10% span, 0–20 mA ±0.03% span
Output
Measurement only; loop power supply 24 Vdc / 35 mA max (non-I.S.)
Supply
(4) AA batteries or USB / universal AC adapter 100–240 Vac 50/60 Hz
Process connection
Via interchangeable Quick-Select AM2 pressure modules (module-dependent)
Housing
High-impact PC-ABS
Download datasheet PDF

Fits portable pressure and loop calibration where you need an Ashcroft ATE-2—handheld calibrator with two hot-swap AM2 module bays covering about 0.25 in H2O to 10,000 psi, 0–20 mA / 0–30 VDC measurement, optional 24 V loop power, logging to 16,000 records, and IP65/NEMA 4X enclosure.

Key advantages

  • Modules — Two hot-swap AM2 bays; 0.25 in H2O–10,000 psi
  • Electrical — 0–20 mA / 0–30 VDC; 24 V loop (non-IS)
  • Logging — Up to 16,000 records; USB + SD
  • Enclosure — IP65/NEMA 4X
  • Display — Up to three simultaneous measurements

Typical applications

  • Field pressure transmitter cal.
  • Switch testing.
  • Multi-parameter logging.

Selection guide

  • AM2 ranges
  • ST vs IS base
  • Temperature modules

Common mistakes

  • Hot-swap without care.
  • IS unit expecting loop power.
  • Ignoring warm-up.

When to RFQ / what to tell us

  1. Base + modules.
  2. Quantity, ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU42206
BrandAshcroft
ModelATE-2
Product typeHandheld Calibrator
TypeHandheld multifunction calibrator (ATE-2)
ModulesTwo hot-swap AM2 bays; pressure 0.25 in H2O–10,000 psi
Electrical0–20 mA / 0–30 VDC; 24 V loop (non-IS)
LoggingUp to 16,000 records; USB + SD
EnclosureIP65/NEMA 4X; optional IS versions

Documents

FAQ

Q: Pressure span? A: Modules from 0.25 in H2O up to 10,000 psi.

Q: Logging? A: Up to 16,000 records.

Q: What should I send for a quote? A: Base/modules, quantity, ship-to.

Links

ParameterValue
Measuring rangePressure modules 0.25 in H2O to 10,000 psi; electrical 0–20 mA or 0–30 Vdc
AccuracyPressure modules ±0.025% to ±0.1% of span; electrical 0–10 Vdc ±0.025% span, 10–30 Vdc ±0.10% span, 0–20 mA ±0.03% span
OutputMeasurement only; loop power supply 24 Vdc / 35 mA max (non-I.S.)
Supply(4) AA batteries or USB / universal AC adapter 100–240 Vac 50/60 Hz
Process connectionVia interchangeable Quick-Select AM2 pressure modules (module-dependent)
HousingHigh-impact PC-ABS
ProtectionIP65 / NEMA 4X
Ambient temperatureOperating −4 to 120 °F (−20 to 49 °C); storage −4 to 158 °F (−20 to 70 °C)
DisplayLCD 2.5 in W × 1.5 in H; 3 simultaneous measurements; backlight; flip screen
CommunicationUSB Micro-B; SD memory card slot
ApprovalsCE; UKCA; FCC CFR47; UL 61010-1; Optional FM/CSA/ATEX IS
Resolution±0.001% span; 99,999 counts
Data LoggingUp to 64 logs / 16,000 records internal; intervals 0.1 s to 24 h
Battery LifeUp to 40 hours with 2 modules installed
Dimensions8.7 in L × 5.1 in W × 3.8 in H
Weight2.4 lb (no modules)