Image unavailable
Image unavailable
Mensor
Mensor CPR4000 Pressure Sensor
Quote on request
No online payment · add lines to RFQ BOM
Mensor CPR4000: reference transducer; 0.02% FS; 0.008% precision; to 210 bar. RFQ.
Download datasheet PDF
RFQ cart holds BOM lines only. Submit for engineering quote.
Key parameters
Accuracy
0.02% FS standard; optional 0.02% IS-50
Type
Reference pressure transducer module CPR4000 (used in CPC4000 industrial pressure controller)
Precision
0.008% FS
Gauge Range
0 ... 0.35 to 0 ... 210 bar (0 ... 5 to 0 ... 3,045 psi)
Bidirectional
-0.17 ... 0.17 to -1 ... 210 bar (-2.5 ... 2.5 to -15 ... 3,045 psi)
Absolute
0 ... 1 to 0 ... 211 bar abs. (0 ... 15 to 0 ... 3,060 psi abs.)
Fits reference pressure transduction where you need a Mensor CPR4000. Reference pressure transducer module used in the CPC4000 industrial pressure controller, with accuracy 0.02% FS standard (optional 0.02% IS-50), precision 0.008% FS, gauge ranges 0…0.35 to 0…210 bar, bidirectional and absolute ranges available, and 365-day calibration interval.
Key advantages
- Accuracy — 0.02% FS standard
- Precision — 0.008% FS
- Gauge — To 210 bar class
- Interval — 365-day calibration
- Host — CPC4000 controller module
Typical applications
- Reference pressure control.
- CPC4000 transducer slots.
- Lab/industrial pressure standards.
Selection guide
- Gauge / bi / absolute
- Accuracy option
- Pressure unit set
Common mistakes
- Wrong range type.
- Ignoring calibration interval.
- Mixing IS-50 vs standard claim.
When to RFQ / what to tell us
- Range/type/options.
- Quantity, ship-to.
Request a Quote · Get selection help
Specs
| Spec | Value |
|---|---|
| SKU | 43064 |
| Brand | Mensor |
| Model | CPR4000 Pressure Sensor |
| Product type | Pressure Sensor |
| Type | Reference pressure transducer module CPR4000 (used in CPC4000 industrial pressure controller) |
| Accuracy | 0.02% FS standard; optional 0.02% IS-50 |
| Precision | 0.008% FS |
| Gauge range | 0 ... 0.35 to 0 ... 210 bar (0 ... 5 to 0 ... 3,045 psi) |
| Bidirectional | -0.17 ... 0.17 to -1 ... 210 bar (-2.5 ... 2.5 to -15 ... 3,045 psi) |
| Absolute | 0 ... 1 to 0 ... 211 bar abs. (0 ... 15 to 0 ... 3,060 psi abs.) |
| Calibration interval | 365 days |
| Pressure units | bar, mbar, psi, MPa, kPa, inHg, inH2O (plus more in controller) |
Documents
FAQ
Q: Accuracy? A: 0.02% FS standard.
Q: Precision? A: 0.008% FS.
Q: What should I send for a quote? A: Range/type/options, quantity, ship-to.
Links
| Parameter | Value |
|---|---|
| Accuracy | 0.02% FS standard; optional 0.02% IS-50 |
| Type | Reference pressure transducer module CPR4000 (used in CPC4000 industrial pressure controller) |
| Precision | 0.008% FS |
| Gauge Range | 0 ... 0.35 to 0 ... 210 bar (0 ... 5 to 0 ... 3,045 psi) |
| Bidirectional | -0.17 ... 0.17 to -1 ... 210 bar (-2.5 ... 2.5 to -15 ... 3,045 psi) |
| Absolute | 0 ... 1 to 0 ... 211 bar abs. (0 ... 15 to 0 ... 3,060 psi abs.) |
| Calibration Interval | 365 days |
| Pressure Units | bar, mbar, psi, MPa, kPa, inHg, inH2O (plus more in controller) |
| Document | WIKA/Mensor data sheet CT 27.40 CPC4000 (CPR4000 transducer section) |