PIE Practical Instrument Electronics

PIE 525B Thermocouple & RTD Calibrator

SKU · 12357 · 525B Thermocouple & RTD Calibrator · Brand: PIE Practical Instrument Electronics
Available
Quote on request
No online payment · add lines to RFQ BOM

Automated TC/RTD calibrator. Source & read; EZ-CHECK. RFQ + selection help.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Functions
Source & read thermocouple, RTD, ohms, millivolts
Thermocouple Types
J K T E R S B N G C D L(J-DIN) U(T-DIN) P(Platinel II); −13.0000 to +80.0000 mV
Rtd Types
Pt 10/50/100/200/500/1000 Ω (3850); Pt100 3902/3916/3926; Cu10/50; Ni120; 0–400.000 / 0–4000.00 Ω
Tc Accuracy
±(0.008% of setting/reading + 0.006 mV)
Cold Junction
±0.05 °C (±0.09 °F); NIST-traceable thermistor
Rtd Accuracy 3 4w
±(0.015% of setting/reading + 0.05 Ω) from 1–10.2 mA excitation
Download datasheet PDF

Fits temperature transmitter and instrument calibration where you need a PIE 525B to source and read thermocouple, RTD, ohms, and millivolts with automated steps.

Key advantages

  • TC + RTD source/read — Source & read thermocouple, RTD, ohms, millivolts
  • Thermocouples — J K T E R S B N G C D L U P; −13.0000 to +80.0000 mV
  • RTDs — Pt 10…1000 Ω (3850) plus Pt100 variants, Cu, Ni; automatic 2/3/4-wire detection
  • Accuracy — TC ±(0.008% + 0.006 mV); CJC ±0.05 °C; RTD 3/4-wire ±(0.015% + 0.05 Ω)
  • Automation — Auto-step 2/3/5/11 or ramp; EZ-CHECK three presets; 0.1°/0.01° selectable

Typical applications

  • Thermocouple and RTD transmitter calibration.
  • Automated step/ramp checks with EZ-CHECK presets.
  • Field temperature loops needing source and read in one tool.

Selection guide

  • Sensor types needed — TC list and RTD curves
  • Wiring — 2/3/4-wire RTD (ConnectSense automatic detection)
  • Step/ramp timing — 5–60 s step times

Common mistakes

  • Treating 525B as a pressure calibrator — it is TC/RTD/Ω/mV.
  • Ignoring cold-junction uncertainty on TC work.
  • Using 2-wire RTD without accounting for lead resistance when 3/4-wire is available.

When to RFQ / what to tell us

  1. TC/RTD types used on site.
  2. Quantity; ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU12357
BrandPIE
Model525B Thermocouple & RTD Calibrator
Product typeTemperature Calibrator
FunctionsSource & read thermocouple, RTD, ohms, millivolts
Thermocouple typesJ K T E R S B N G C D L(J-DIN) U(T-DIN) P(Platinel II); −13.0000 to +80.0000 mV
RTD typesPt 10/50/100/200/500/1000 Ω (3850); Pt100 3902/3916/3926; Cu10/50; Ni120; 0–400.000 / 0–4000.00 Ω
TC accuracy±(0.008% of setting/reading + 0.006 mV)
Cold junction±0.05 °C (±0.09 °F); NIST-traceable thermistor
RTD accuracy (3/4-wire)±(0.015% of setting/reading + 0.05 Ω) from 1–10.2 mA excitation
Resolution0.1° and 0.01° F/C selectable
Auto step2/3/5/11 steps or ramp; 5–60 s step times; EZ-CHECK three presets

Documents

FAQ

Q: What does 525B do? A: Source and read thermocouple, RTD, ohms, and millivolts.

Q: RTD wiring? A: Automatic 2/3/4-wire detection (ConnectSense).

Q: What should I send for a quote? A: Sensor types needed, quantity, ship-to.

Links

ParameterValue
FunctionsSource & read thermocouple, RTD, ohms, millivolts
Thermocouple TypesJ K T E R S B N G C D L(J-DIN) U(T-DIN) P(Platinel II); −13.0000 to +80.0000 mV
Rtd TypesPt 10/50/100/200/500/1000 Ω (3850); Pt100 3902/3916/3926; Cu10/50; Ni120; 0–400.000 / 0–4000.00 Ω
Tc Accuracy±(0.008% of setting/reading + 0.006 mV)
Cold Junction±0.05 °C (±0.09 °F); NIST-traceable thermistor
Rtd Accuracy 3 4w±(0.015% of setting/reading + 0.05 Ω) from 1–10.2 mA excitation
Resolution0.1° and 0.01° F/C selectable
Auto Step2/3/5/11 steps or ramp; 5–60 s step times; EZ-CHECK three presets
Rtd ConnectsenseAutomatic 2/3/4-wire detection
Power SizeFour AA; ~50 h; 143×76×41 mm; 0.34 kg with boot