PIE Practical Instrument Electronics
PIE 525B Thermocouple & RTD Calibrator
Quote on request
No online payment · add lines to RFQ BOM
Automated TC/RTD calibrator. Source & read; EZ-CHECK. RFQ + selection help.
Download datasheet PDF
RFQ cart holds BOM lines only. Submit for engineering quote.
Key parameters
Functions
Source & read thermocouple, RTD, ohms, millivolts
Thermocouple Types
J K T E R S B N G C D L(J-DIN) U(T-DIN) P(Platinel II); −13.0000 to +80.0000 mV
Rtd Types
Pt 10/50/100/200/500/1000 Ω (3850); Pt100 3902/3916/3926; Cu10/50; Ni120; 0–400.000 / 0–4000.00 Ω
Tc Accuracy
±(0.008% of setting/reading + 0.006 mV)
Cold Junction
±0.05 °C (±0.09 °F); NIST-traceable thermistor
Rtd Accuracy 3 4w
±(0.015% of setting/reading + 0.05 Ω) from 1–10.2 mA excitation
Fits temperature transmitter and instrument calibration where you need a PIE 525B to source and read thermocouple, RTD, ohms, and millivolts with automated steps.
Key advantages
- TC + RTD source/read — Source & read thermocouple, RTD, ohms, millivolts
- Thermocouples — J K T E R S B N G C D L U P; −13.0000 to +80.0000 mV
- RTDs — Pt 10…1000 Ω (3850) plus Pt100 variants, Cu, Ni; automatic 2/3/4-wire detection
- Accuracy — TC ±(0.008% + 0.006 mV); CJC ±0.05 °C; RTD 3/4-wire ±(0.015% + 0.05 Ω)
- Automation — Auto-step 2/3/5/11 or ramp; EZ-CHECK three presets; 0.1°/0.01° selectable
Typical applications
- Thermocouple and RTD transmitter calibration.
- Automated step/ramp checks with EZ-CHECK presets.
- Field temperature loops needing source and read in one tool.
Selection guide
- Sensor types needed — TC list and RTD curves
- Wiring — 2/3/4-wire RTD (ConnectSense automatic detection)
- Step/ramp timing — 5–60 s step times
Common mistakes
- Treating 525B as a pressure calibrator — it is TC/RTD/Ω/mV.
- Ignoring cold-junction uncertainty on TC work.
- Using 2-wire RTD without accounting for lead resistance when 3/4-wire is available.
When to RFQ / what to tell us
- TC/RTD types used on site.
- Quantity; ship-to.
Request a Quote · Get selection help
Specs
| Spec | Value |
|---|---|
| SKU | 12357 |
| Brand | PIE |
| Model | 525B Thermocouple & RTD Calibrator |
| Product type | Temperature Calibrator |
| Functions | Source & read thermocouple, RTD, ohms, millivolts |
| Thermocouple types | J K T E R S B N G C D L(J-DIN) U(T-DIN) P(Platinel II); −13.0000 to +80.0000 mV |
| RTD types | Pt 10/50/100/200/500/1000 Ω (3850); Pt100 3902/3916/3926; Cu10/50; Ni120; 0–400.000 / 0–4000.00 Ω |
| TC accuracy | ±(0.008% of setting/reading + 0.006 mV) |
| Cold junction | ±0.05 °C (±0.09 °F); NIST-traceable thermistor |
| RTD accuracy (3/4-wire) | ±(0.015% of setting/reading + 0.05 Ω) from 1–10.2 mA excitation |
| Resolution | 0.1° and 0.01° F/C selectable |
| Auto step | 2/3/5/11 steps or ramp; 5–60 s step times; EZ-CHECK three presets |
Documents
FAQ
Q: What does 525B do? A: Source and read thermocouple, RTD, ohms, and millivolts.
Q: RTD wiring? A: Automatic 2/3/4-wire detection (ConnectSense).
Q: What should I send for a quote? A: Sensor types needed, quantity, ship-to.
Links
| Parameter | Value |
|---|---|
| Functions | Source & read thermocouple, RTD, ohms, millivolts |
| Thermocouple Types | J K T E R S B N G C D L(J-DIN) U(T-DIN) P(Platinel II); −13.0000 to +80.0000 mV |
| Rtd Types | Pt 10/50/100/200/500/1000 Ω (3850); Pt100 3902/3916/3926; Cu10/50; Ni120; 0–400.000 / 0–4000.00 Ω |
| Tc Accuracy | ±(0.008% of setting/reading + 0.006 mV) |
| Cold Junction | ±0.05 °C (±0.09 °F); NIST-traceable thermistor |
| Rtd Accuracy 3 4w | ±(0.015% of setting/reading + 0.05 Ω) from 1–10.2 mA excitation |
| Resolution | 0.1° and 0.01° F/C selectable |
| Auto Step | 2/3/5/11 steps or ramp; 5–60 s step times; EZ-CHECK three presets |
| Rtd Connectsense | Automatic 2/3/4-wire detection |
| Power Size | Four AA; ~50 h; 143×76×41 mm; 0.34 kg with boot |