Bently Nevada
Bently Nevada Model 9100D Portable Vibration Calibrator
Quote on request
No online payment · add lines to RFQ BOM
BN/TMS 9100D: portable shaker; frequency/amplitude cal for vibration sensors. RFQ.
Download datasheet PDF
RFQ cart holds BOM lines only. Submit for engineering quote.
Key parameters
Frequency Range
5 Hz to 10 kHz (100 g payload; 300–600 k CPM)
Max Amplitude 10g Payload
20 g pk / 196 m/s² pk; 20 in/s pk; 500 mm/s pk; 150 mils pk-pk @ 50 Hz
Max Amplitude 500g Payload
2.5 g pk / 24.5 m/s² pk; 3.5 in/s pk; 90 mm/s pk @ 50 Hz
Maximum Payload
800 g (operating range reduced at higher payloads)
Sensor Mount
1/4-28 thread mounting platform
Test Operation
Manual (closed loop) or semi-automatic; auto-payload via reference accelerometer
Fits field vibration sensor checks where you need a Bently Nevada / TMS Model 9100D—portable shaker table for calibrating accelerometers and velocity sensors across a defined frequency and amplitude envelope with onboard readout accuracy per OEM specifications.
Key advantages
- Form — Portable vibration shaker / calibrator
- Use — Sensor amplitude/frequency checks
- Payload — Defined max payload per OEM
- Readout — Onboard amplitude accuracy class
- Platform — Bently Nevada / TMS 9100D
Typical applications
- Accelerometer field cal.
- Velocity sensor checks.
- Route-based vibration QA.
Selection guide
- Payload vs amplitude
- Sensor mount adapter
- Frequency band
Common mistakes
- Overloading payload.
- Loose mount stud.
- Ignoring warm-up.
When to RFQ / what to tell us
- Adapters/options.
- Quantity, ship-to.
Request a Quote · Get selection help
Specs
| Spec | Value |
|---|---|
| SKU | 43971 |
| Brand | Bently Nevada |
| Model | Model 9100D Shaker Table |
| Product type | Vibration Calibrator |
| Frequency range | 5 Hz to 10 kHz (100 g payload; 300–600 k CPM) |
| Max amplitude 10g payload | 20 g pk / 196 m/s² pk; 20 in/s pk; 500 mm/s pk; 150 mils pk-pk @ 50 Hz |
| Max amplitude 500g payload | 2.5 g pk / 24.5 m/s² pk; 3.5 in/s pk; 90 mm/s pk @ 50 Hz |
| Maximum payload | 800 g (operating range reduced at higher payloads) |
| Sensor mount | 1/4-28 thread mounting platform |
| Readout accuracy amplitude | <1% up to 10 g pk (per datasheet accuracy of readout) |
| Dimensions | 8.5 × 12 × 10 in (22 × 30.5 × 28 cm) |
| Weight | 18 lb (8.2 kg) |
Documents
FAQ
Q: Platform? A: Bently Nevada / TMS Model 9100D.
Q: Use? A: Portable vibration sensor calibration.
Q: What should I send for a quote? A: Adapters/options, quantity, ship-to.
Links
| Parameter | Value |
|---|---|
| Frequency Range | 5 Hz to 10 kHz (100 g payload; 300–600 k CPM) |
| Max Amplitude 10g Payload | 20 g pk / 196 m/s² pk; 20 in/s pk; 500 mm/s pk; 150 mils pk-pk @ 50 Hz |
| Max Amplitude 500g Payload | 2.5 g pk / 24.5 m/s² pk; 3.5 in/s pk; 90 mm/s pk @ 50 Hz |
| Maximum Payload | 800 g (operating range reduced at higher payloads) |
| Sensor Mount | 1/4-28 thread mounting platform |
| Test Operation | Manual (closed loop) or semi-automatic; auto-payload via reference accelerometer |
| Memory | 500 calibration records; 30 data points per record; model/serial/orientation/notes |
| Readout Accuracy Amplitude | <1% up to 10 g pk (per datasheet accuracy of readout) |
| Dimensions | 8.5 × 12 × 10 in (22 × 30.5 × 28 cm) |
| Weight | 18 lb (8.2 kg) |
| Operating Temperature | 32 to 122°F (0 to 50°C) |
| Battery | 12 Vdc 4 Ah sealed gel lead-acid; ~18 h @ 100 Hz / 1 g pk / 100 g payload (new) |
| Reference | Built-in NIST-traceable reference accelerometer permanently attached to armature |
| Programmability | Up to 30 test points per routine with Pass/Fail bounds |