Endress+Hauser

Endress+Hauser Memosens CPS11E Orbisint pH Sensor

SKU · 36945 · Memosens CPS11E · Brand: Endress+Hauser
Available
Quote on request
No online payment · add lines to RFQ BOM

CPS11E: pH 0–14 apps; to 135 °C; PTFE junction; Memosens 2.0. RFQ app.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Measuring range
App A pH 1–12; App B pH 0–14; App F pH 0–10
Approvals
ATEX/IECEx/CSA C/US/NEPSI/Japan Ex/INMETRO Zone 0/1/2 options
Type
Digital Memosens 2.0 pH sensor for standard process/environmental applications (CPS11E Orbisint)
Process Temperature
A: −15…80 °C; B: 0…135 °C; F: 0…70 °C
Process Pressure
A/B: 0.8…17 bar abs; F: 0.8…7 bar abs
Junction
Dirt-repellent PTFE ring junction; gel compact electrode
Download datasheet PDF

Fits standard process and environmental pH where you need an Endress+Hauser Memosens CPS11E Orbisint—application ranges covering pH 0–14, PTFE ring junction, integrated NTC 30K, and inductive Memosens 2.0 connection.

Key advantages

  • Range — App A 1–12; App B 0–14; App F 0–10
  • Temperature — A −15…80 °C; B 0…135 °C; F 0…70 °C
  • Pressure — A/B 0.8…17 bar abs; F 0.8…7 bar abs
  • Junction — Dirt-repellent PTFE ring; gel compact
  • Connection — Inductive Memosens 2.0; shaft lengths 120–425 mm

Typical applications

  • Process water pH.
  • Wastewater monitoring.
  • General chemical pH with lab-calibrated swap.

Selection guide

  • Application code A/B/F
  • Shaft length
  • Transmitter (CM42/CM44x)

Common mistakes

  • Wrong application code for temperature.
  • Exceeding pressure rating.
  • Analog cable on Memosens head.

When to RFQ / what to tell us

  1. App code, length.
  2. Transmitter, quantity, ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU36945
BrandEndress+Hauser
ModelMemosens CPS11E Orbisint pH Sensor
Product typepH Sensor
TypeDigital Memosens 2.0 pH sensor for standard process/environmental applications (CPS11E Orbisint)
Measuring rangeApp A pH 1–12; App B pH 0–14; App F pH 0–10
Process temperatureA: −15…80 °C; B: 0…135 °C; F: 0…70 °C
Process pressureA/B: 0.8…17 bar abs; F: 0.8…7 bar abs
JunctionDirt-repellent PTFE ring junction; gel compact electrode
Temperature sensorIntegrated NTC 30K
ConnectionInductive digital Memosens 2.0 plug-in head

Documents

FAQ

Q: Memosens version? A: Memosens 2.0 inductive plug-in head.

Q: Max process T? A: Up to 135 °C on application B.

Q: What should I send for a quote? A: App/length, quantity, ship-to.

Links

ParameterValue
Measuring rangeApp A pH 1–12; App B pH 0–14; App F pH 0–10
ApprovalsATEX/IECEx/CSA C/US/NEPSI/Japan Ex/INMETRO Zone 0/1/2 options
TypeDigital Memosens 2.0 pH sensor for standard process/environmental applications (CPS11E Orbisint)
Process TemperatureA: −15…80 °C; B: 0…135 °C; F: 0…70 °C
Process PressureA/B: 0.8…17 bar abs; F: 0.8…7 bar abs
JunctionDirt-repellent PTFE ring junction; gel compact electrode
Temperature SensorIntegrated NTC 30K
ShaftGlass shaft Ø12 mm; lengths 120/225/360/425 mm
ConnectionInductive digital Memosens 2.0 plug-in head
Ingress ProtectionIP68
ApplicationLong-term monitoring in stable processes; water & wastewater; chemicals; power
DocumentTI01493C/07/EN/03.25-00