Rosemount

Rosemount 396P / 396PVP TUpH pH/ORP Sensors

SKU · 36951 · 396P / 396PVP · Brand: Rosemount
Available
Quote on request
No online payment · add lines to RFQ BOM

396P/PVP: pH 0–14; ORP ±1500 mV; to 100 °C / 150 psig; TUpH. RFQ.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Measuring range
pH 0–14; ORP −1500 to +1500 mV (electrode options)
Process connection
1 in (25 mm) MNPT front and rear facing threads (insertion/submersion/flow-through)
Type
TUpH pH/ORP sensor for dirty/abrasive/coating/high-solids service (Rosemount 396P / 396PVP)
Process Temperature
Max 212 °F (100 °C)
Process Pressure
Max 150 psig (10.34 barg)
Body Material
Reinforced polypropylene
Download datasheet PDF

Fits dirty, abrasive, coating, and high-solids pH/ORP service where you need a Rosemount 396P or 396PVP TUpH sensor—pH 0–14 or ORP −1500…+1500 mV, reinforced polypropylene body, and large-area TUpH junction.

Key advantages

  • pH / ORP — 0–14 pH; −1500…+1500 mV
  • Process — Max 100 °C; max 150 psig
  • Body — Reinforced polypropylene; 1 in MNPT front and rear
  • Reference — TUpH large-area junction; helical poisoning-resistant path
  • Electrodes — GPLR hemi glass; optional flat pH; platinum ORP

Typical applications

  • High-solids wastewater pH.
  • Coating / abrasive slurry service.
  • Insertion, submersion, or flow-through.

Selection guide

  • 396P vs 396PVP connector
  • Glass style / ORP
  • Mounting method

Common mistakes

  • Exceeding 150 psig / 100 °C.
  • Using standard junction in coating service.
  • Wrong cable for VP variant.

When to RFQ / what to tell us

  1. P vs PVP, electrode.
  2. Quantity, ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU36951
BrandRosemount
ModelModel 396P / 396PVP TUpH pH/ORP Sensors
Product typepH ORP Sensor
TypeTUpH pH/ORP sensor for dirty/abrasive/coating/high-solids service (Rosemount 396P / 396PVP)
Measuring rangepH 0–14; ORP −1500 to +1500 mV (electrode options)
Process temperatureMax 212 °F (100 °C)
Process pressureMax 150 psig (10.34 barg)
Process connection1 in (25 mm) MNPT front and rear facing threads (insertion/submersion/flow-through)
Body materialReinforced polypropylene
ReferenceTUpH large-area junction; poisoning-resistant helical reference pathway

Documents

FAQ

Q: Max temperature? A: 100 °C (212 °F).

Q: Process connection? A: 1 in MNPT front and rear facing threads.

Q: What should I send for a quote? A: Variant/electrode, quantity, ship-to.

Links

ParameterValue
Measuring rangepH 0–14; ORP −1500 to +1500 mV (electrode options)
Process connection1 in (25 mm) MNPT front and rear facing threads (insertion/submersion/flow-through)
TypeTUpH pH/ORP sensor for dirty/abrasive/coating/high-solids service (Rosemount 396P / 396PVP)
Process TemperatureMax 212 °F (100 °C)
Process PressureMax 150 psig (10.34 barg)
Body MaterialReinforced polypropylene
ReferenceTUpH large-area junction; poisoning-resistant helical reference pathway
ElectrodesGPLR hemi glass; optional flat pH glass; platinum ORP
Cable396P integral cable (preamp options); 396PVP Variopol VP8 (cable sold separately)
Temperature CompensationPt-100 / options for Rosemount 1056/1066/56/5081 family (code 55) or legacy transmitters (code 54)
SmartSMART enabled preamplifier options for pH
DocumentPDS 00813-0100-3396 Rev AD + QSG