Yokogawa

Rosemount 3300HT / 3300HTVP High-Temp pH Sensor

SKU · 36953 · PROPLUS YPP6060 Process Meter · Brand: Yokogawa
Available
Quote on request
No online payment · add lines to RFQ BOM

3300HT: pH 0–14; to 155 °C; 400 psig; Ti body; PERpH-X. RFQ.

Download datasheet PDF

RFQ cart holds BOM lines only. Submit for engineering quote.

Key parameters

Measuring range
Dual inputs: 0–20 mA, 4–20 mA, 0–5 V, 1–5 V, ±10 V (each channel independently programmed)
Accuracy
±0.03% of calibrated span ±1 count (LIM6060YK); square-root/exponent accuracy 10–100% of span
Output
Optional 2 or 4 SPDT relays; isolated 4–20 mA retransmission
Supply
Universal 85–265 VAC or 12/24 VDC models
Display
Dual-line 6-digit LED, 0.60 in / 0.46 in; standard or SunBright outdoor
Communication
USB on-board; optional RS-232 / RS-485 Modbus RTU
Download datasheet PDF

Fits high-temperature and high-pressure pH/ORP where you need a Rosemount 3300HT or 3300HTVP PERpH-X sensor—pH 0–14, ORP −1500…+1500 mV, titanium body, rebuildable double-junction reference, and up to 155 °C / 400 psig class.

Key advantages

  • Range — pH 0–14; ORP −1500…+1500 mV
  • Temperature — 5–155 °C without preamp; to 100 °C with preamp
  • Pressure — Max 400 psig
  • Body — Titanium; PTFE junction; Pt-100
  • Cable — Integral (HT) or VP8 (HTVP)

Typical applications

  • High-temperature process pH.
  • Retractable insertion assemblies.
  • Poisoning / fouling resistant electrolyte options.

Selection guide

  • HT vs HTVP connector
  • Electrolyte option
  • Process connector / length

Common mistakes

  • Using without required 1 in compression connector.
  • Exceeding preamp temperature limit.
  • Wrong O-ring material for chemistry.

When to RFQ / what to tell us

  1. HT vs HTVP, electrolyte.
  2. Quantity, ship-to.

Request a Quote · Get selection help

Specs

SpecValue
SKU36953
BrandRosemount
ModelModel 3300HT / 3300HTVP pH Sensor
Product typepH ORP Sensor
TypePERpH-X high-performance pH/ORP sensor (Rosemount 3300HT / 3300HTVP)
pH range0–14 pH
ORP range−1500 to +1500 mV
Temperature5–155 °C without preamp; to 100 °C with preamp
PressureMax 400 psig (2859 kPa)
BodyTitanium; PTFE junction; Pt-100 RTD
ReferenceDouble junction; replaceable process-side electrolyte
Connection1 in compression process connector; integral cable (HT) or VP8 (HTVP)

Documents

FAQ

Q: Max temperature? A: 155 °C without preamplifier.

Q: Max pressure? A: 400 psig.

Q: What should I send for a quote? A: Variant/electrolyte, quantity, ship-to.

Links

ParameterValue
Measuring rangeDual inputs: 0–20 mA, 4–20 mA, 0–5 V, 1–5 V, ±10 V (each channel independently programmed)
Accuracy±0.03% of calibrated span ±1 count (LIM6060YK); square-root/exponent accuracy 10–100% of span
OutputOptional 2 or 4 SPDT relays; isolated 4–20 mA retransmission
SupplyUniversal 85–265 VAC or 12/24 VDC models
DisplayDual-line 6-digit LED, 0.60 in / 0.46 in; standard or SunBright outdoor
CommunicationUSB on-board; optional RS-232 / RS-485 Modbus RTU
FunctionsMath: addition, difference, absolute difference, average, weighted average, multiplication, division, min/max, draw, ratio, concentration; 32-point linearization; square-root; exponential; round horizontal tank; multi-pump alternation; tare; max/min
Transmitter PowerIsolated 24 VDC @ 200 mA (VAC models) / 100 mA (VDC models)
Enclosure1/8 DIN panel meter; NEMA 4X, IP65 front
Operating Temperature−40 to +65 °C (−40 to +149 °F)
Form Factor1/8 DIN dual-line digital panel meter
Linearization32-point, square root, or programmable exponent per input channel