Parámetros clave

Accuracy
±0.5 °C
Supply
100–130 or 208–240 VAC (specify)
Notes
Specs from Isotech QuickCal Blackbody 2025 datasheet. No Instrumart.
Temperature Range
30 °C to 350 °C
Stability
±0.2 °C
Display Resolution
0.1 °C
Descargar datasheet PDF
ParámetroValor
Accuracy±0.5 °C
Supply100–130 or 208–240 VAC (specify)
NotesSpecs from Isotech QuickCal Blackbody 2025 datasheet. No Instrumart.
Temperature Range30 °C to 350 °C
Stability±0.2 °C
Display Resolution0.1 °C
Emissivity>0.99 overall (surface coating 0.98; cavity geometry raises effective emissivity)
Target Size25 × 115 mm cylindrical cavity with end cone
Heating Time9 minutes
Power300 W
DimensionsH 65 × W 152 × D 175 mm
Weight1.5 kg
How To OrderModel 550-02 Blackbody Source; state voltage required