Endress+Hauser

Endress+Hauser Indumax CLS50D Conductivity Sensor

SKU · 47594 · Indumax CLS50D Conductivity Sensor · Marca: Endress+Hauser
Disponible
Cotización a solicitud
Sin pago en línea · líneas en BOM RFQ
Descargar datasheet PDF

RFQ solo líneas BOM. Envíe para cotización de ingeniería.

Parámetros clave

Communication
CLS50D Memosens digital to Liquiline CM44x/R or CM42; CLS50 analog to CM42 / Liquisys CLM223/253
Measuring Principle
Inductive (toroidal) conductivity
Conductivity Range
2 μS/cm to 2000 mS/cm (uncompensated)
Temperature Sensor
Integrated coated Pt100, class A
Process Temperature
CLS50D PEEK versions typically −20 to 125 °C (−4 to 260 °F); CLS50 PEEK up to 180 °C (356 °F) on selected versions
Process Pressure
Max 41 bar (595 psi) absolute depending on version (see T-p chart)
Descargar datasheet PDF
ParámetroValor
CommunicationCLS50D Memosens digital to Liquiline CM44x/R or CM42; CLS50 analog to CM42 / Liquisys CLM223/253
Measuring PrincipleInductive (toroidal) conductivity
Conductivity Range2 μS/cm to 2000 mS/cm (uncompensated)
Temperature SensorIntegrated coated Pt100, class A
Process TemperatureCLS50D PEEK versions typically −20 to 125 °C (−4 to 260 °F); CLS50 PEEK up to 180 °C (356 °F) on selected versions
Process PressureMax 41 bar (595 psi) absolute depending on version (see T-p chart)
Ingress ProtectionIP68 / NEMA Type 6 (installed with genuine seal)
MaterialsPFA or PEEK wetted sensor; adapters PEEK / 1.4571 per order code
CableFixed cable; extend with CYK11 (CLS50D) or CLK6 (CLS50); total length up to 55–100 m class per TI
Response TimeConductivity t95 ≤ 2 s; temperature t90 ≤ 7 min (PEEK)
Ex ApprovalEx ia IIC T4/T6 options
DocumentTI00182C/07/EN/22.26-00; BA00182CEN