Parámetros clave
Measuring range
Integral orifice primary for 0.5–1.5 in. (15–40 mm) lines; bore sizes from ~0.010 in. upward per ordering table
Accuracy
Up to ±1.00% with precision honed pipe sections; self-centering plate reduces small-line installation error
Output
Primary element DP; assemble-to Rosemount DP transmitters (e.g. 3051SFP integrated meter)
Process connection
Meter-run flanged / threaded process connections per model; direct or remote transmitter mount
Wetted parts
Orifice plate and body materials per PDS (SST typical)
Media temperature
Direct mount −40 to 450 °F (−40 to 232 °C); remote mount −112 to 850 °F (−80 to 454 °C) per Integral Orifice RM
| Parámetro | Valor |
|---|---|
| Measuring range | Integral orifice primary for 0.5–1.5 in. (15–40 mm) lines; bore sizes from ~0.010 in. upward per ordering table |
| Accuracy | Up to ±1.00% with precision honed pipe sections; self-centering plate reduces small-line installation error |
| Output | Primary element DP; assemble-to Rosemount DP transmitters (e.g. 3051SFP integrated meter) |
| Process connection | Meter-run flanged / threaded process connections per model; direct or remote transmitter mount |
| Wetted parts | Orifice plate and body materials per PDS (SST typical) |
| Media temperature | Direct mount −40 to 450 °F (−40 to 232 °C); remote mount −112 to 850 °F (−80 to 454 °C) per Integral Orifice RM |
| Ambient temperature | Per assembled transmitter ambient limits |
| Communication | Via assembled DP transmitter (HART / Fieldbus / WirelessHART) |
| Approvals | Hazardous location / product certifications per assembled transmitter and 1195 docs |
| Notes | Specs from Emerson DP Flow PDS 00813-0100-4485 (1195 section) and Integral Orifice RM 00809-0100-4686. |