Parámetros clave

Measuring range
0.25 inH2O to 10,000 psi via permanently installed PPT-1/PPT-2 pressure modules (single or dual)
Accuracy
From ±0.1% to ±0.025% of span depending on range/module (low-pressure capacitive ±0.1%/±0.06%/±0.07%; medium/high ±0.025%/±0.05%/±0.1%)
Output
RS-232 digital interface; dual-line LCD display (no analog process output)
Supply
110 Vac/60 Hz adapter standard; optional 220 Vac/50 Hz, 100 Vac, NiCd or alkaline pack
Process connection
1/8 NPT female
Wetted parts
PPM1 non-isolated (clean dry gas); PPM2 316 stainless steel media compatible
Descargar datasheet PDF
ParámetroValor
Measuring range0.25 inH2O to 10,000 psi via permanently installed PPT-1/PPT-2 pressure modules (single or dual)
AccuracyFrom ±0.1% to ±0.025% of span depending on range/module (low-pressure capacitive ±0.1%/±0.06%/±0.07%; medium/high ±0.025%/±0.05%/±0.1%)
OutputRS-232 digital interface; dual-line LCD display (no analog process output)
Supply110 Vac/60 Hz adapter standard; optional 220 Vac/50 Hz, 100 Vac, NiCd or alkaline pack
Process connection1/8 NPT female
Wetted partsPPM1 non-isolated (clean dry gas); PPM2 316 stainless steel media compatible
Display2-line LCD, 0.38 in character height, 16 characters/line; 100 ms update
ApprovalsNIST-traceable calibration certificate included per pressure module
NotesSpecs from Ashcroft PT-1 datasheet Rev E (2023-03-19) and PT Digital Indicator operating manual. Bench-top secondary transfer standard.
Resolution±0.002% of span (max.)
Repeatability±0.01% of span (psi ranges and 1–200 inH2O); ±0.02% of span below 1 inH2O
Operating Temperature32 to 120 °F (0 to 49 °C); compensated 20 to 120 °F
Case MaterialHigh-impact ABS, black