Image unavailable
Image unavailable
Endress+Hauser
Endress+Hauser Smartec CLD18 Compact Conductivity Device
Цена по запросу
Без онлайн-оплаты · строки в RFQ BOM
Скачать datasheet PDF
RFQ только BOM. Отправьте для инженерного КП.
Ключевые параметры
Measuring Principle
Inductive (toroidal) conductivity
Conductivity Range
Recommended 200 μS/cm to 1000 mS/cm (uncompensated)
Temperature Range Meas
−10 to 130 °C (14 to 266 °F)
Measured Error
Conductivity ±(2.0% of measured value + 20 μS/cm); temperature ±1.5 K
Repeatability
Conductivity max 0.5% of measured value ±5 μS/cm ±2 digits
Outputs
Conductivity and temperature 0/4–20 mA galvanically isolated; optional IO-Link 1.1.3
| Параметр | Значение |
|---|---|
| Measuring Principle | Inductive (toroidal) conductivity |
| Conductivity Range | Recommended 200 μS/cm to 1000 mS/cm (uncompensated) |
| Temperature Range Meas | −10 to 130 °C (14 to 266 °F) |
| Measured Error | Conductivity ±(2.0% of measured value + 20 μS/cm); temperature ±1.5 K |
| Repeatability | Conductivity max 0.5% of measured value ±5 μS/cm ±2 digits |
| Outputs | Conductivity and temperature 0/4–20 mA galvanically isolated; optional IO-Link 1.1.3 |
| Ingress Protection | IP69 (EN 40050); NEMA Type 6P |
| Response Time | Conductivity t95 < 1.5 s; temperature t90 < 20 s |
| Sensor Material | Food-safe PEEK; hygienic design EHEDG / 3-A class |
| Housing | Stainless steel 1.4308 or plastic PBT GF; high-pressure steam cleanable |
| Weight | Stainless max 1.870 kg; plastic max 1.070 kg |
| Applications | Beverage phase separation, CIP concentration control, industrial water, pickling rinse |
| Document | TI01080C/07/EN/06.21; BA01149CEN |