Rosemount

Rosemount Model 5081 Transmitters

SKU · 30764 · 5081 · Бренд: Rosemount
В наличии
Цена по запросу
Без онлайн-оплаты · строки в RFQ BOM
Скачать datasheet PDF

RFQ только BOM. Отправьте для инженерного КП.

Ключевые параметры

Measuring range
Single-channel: pH/ORP, contacting/toroidal conductivity, oxygen (ppm/ppb), free/total chlorine, ozone (model/sensor dependent)
Accuracy
Analog output accuracy ±0.05 mA; measurement accuracy sensor-dependent (e.g. oxygen 0–99 ppm with 499A DO)
Output
Two-wire 4–20 mA with HART or FOUNDATION Fieldbus
Supply
HART: 12–42.4 Vdc at terminals (30 Vdc max IS); 250 Ω min load for HART
Protection
Type 4X (IP65) cast aluminum epoxy-coated explosion-proof enclosure
Ambient temperature
−4 to 149 °F (−20 to 65 °C); storage −22 to 176 °F (−30 to 80 °C); RH 0–95% non-condensing
Скачать datasheet PDF
ПараметрЗначение
Measuring rangeSingle-channel: pH/ORP, contacting/toroidal conductivity, oxygen (ppm/ppb), free/total chlorine, ozone (model/sensor dependent)
AccuracyAnalog output accuracy ±0.05 mA; measurement accuracy sensor-dependent (e.g. oxygen 0–99 ppm with 499A DO)
OutputTwo-wire 4–20 mA with HART or FOUNDATION Fieldbus
SupplyHART: 12–42.4 Vdc at terminals (30 Vdc max IS); 250 Ω min load for HART
ProtectionType 4X (IP65) cast aluminum epoxy-coated explosion-proof enclosure
Ambient temperature−4 to 149 °F (−20 to 65 °C); storage −22 to 176 °F (−30 to 80 °C); RH 0–95% non-condensing
DisplayTwo-line 7-segment LCD (0.8 in process / 0.3 in secondary); rotatable display board
ApprovalsIntrinsically safe; Non-incendive; Explosion-proof hazardous location approvals
MountingExplosion-proof field housing; ¾-in NPT conduit openings
Sensor typeCompatible with most Rosemount Analytical sensors including 499A amperometric DO/chlorine/ozone
NotesSpecs from Emerson Rosemount 5081 PDS 00813-0100-3581 Rev T + RM + emerson.com product page.
Dimensions6.3 × 6.9 × 6.4 in (160.5 × 175.3 × 161.3 mm)
Weight10 lb / shipping 11 lb (4.5 / 5.0 kg)