Endress+Hauser
Endress+Hauser Oxymax COS61D Optical Oxygen Sensor
Цена по запросу
Без онлайн-оплаты · строки в RFQ BOM
Скачать datasheet PDF
RFQ только BOM. Отправьте для инженерного КП.
Ключевые параметры
Measuring range
0 to 20 mg/l; 0 to 400 hPa; 0 to 200 %SAT (20 °C / 1013 hPa reference)
Accuracy
<12 mg/l: 0.01 mg/l or ±1% of reading; 12–20 mg/l: ±2% of reading (COS61D)
Output
Digital Memosens (Liquiline CM44x/CM44xR/CM44P) or fixed-cable/SXP to Liquisys
Wetted parts
Sensor shaft stainless steel 1.4571; membrane cap POM
Protection
IP67/IP68 (per connection/approval)
Approvals
CE; Ex variants per TI
| Параметр | Значение |
|---|---|
| Measuring range | 0 to 20 mg/l; 0 to 400 hPa; 0 to 200 %SAT (20 °C / 1013 hPa reference) |
| Accuracy | <12 mg/l: 0.01 mg/l or ±1% of reading; 12–20 mg/l: ±2% of reading (COS61D) |
| Output | Digital Memosens (Liquiline CM44x/CM44xR/CM44P) or fixed-cable/SXP to Liquisys |
| Wetted parts | Sensor shaft stainless steel 1.4571; membrane cap POM |
| Protection | IP67/IP68 (per connection/approval) |
| Approvals | CE; Ex variants per TI |
| Sensor type | Optical DO via fluorescence quenching; no flow required |
| Notes | Specs from Endress+Hauser TI00387CEN (Oxymax COS61D/COS61) + product page. |
| Process Temperature | −5 to +60 °C (20 to 140 °F) |
| Process Pressure | 1 to 10 bar abs (14.5 to 145 psi) |
| Response Time | t90 ≈ 60 s (air to nitrogen, reference conditions) |