Image unavailable
Image unavailable
Rotronic
Rotronic HygroGen2-S/XL Humidity and Temperature Generator
Цена по запросу
Без онлайн-оплаты · строки в RFQ BOM
Скачать datasheet PDF
RFQ только BOM. Отправьте для инженерного КП.
Ключевые параметры
Approvals
CE EN 61010-1; UKCA; EMC EN 61326-1; FDA 21 CFR Part 11 HW5 software
Humidity Range
5–95 %RH standard; optional extension 2–99 %RH
Temperature Range
0–60 °C standard; optional extension to −5 °C
Stability
<±0.1 %RH and <±0.01 °C
Chamber Volume
HG2-S: 2 L (1.5 L working); HG2-XL: 20 L (17 L working)
Humidity Change Time
5–95 %RH to 0.1 %RH stability: <5 min (S) / <15 min (XL)
| Параметр | Значение |
|---|---|
| Approvals | CE EN 61010-1; UKCA; EMC EN 61326-1; FDA 21 CFR Part 11 HW5 software |
| Humidity Range | 5–95 %RH standard; optional extension 2–99 %RH |
| Temperature Range | 0–60 °C standard; optional extension to −5 °C |
| Stability | <±0.1 %RH and <±0.01 °C |
| Chamber Volume | HG2-S: 2 L (1.5 L working); HG2-XL: 20 L (17 L working) |
| Humidity Change Time | 5–95 %RH to 0.1 %RH stability: <5 min (S) / <15 min (XL) |
| Temperature Change Time | 23–50 °C to 0.01 °C stability: <5 min (S) / <15 min (XL) |
| Typical Calibration Uncertainty | ±1.5 %RH (k=2) at 23 °C; ±0.15 °C (k=2) 15–50 °C |
| Generation Method | Mixed flow with desiccant drier and piezoelectric humidifier; Peltier with mixing fan |
| Probe Mounts | Up to 6 probes (S) or up to 19 (XL) through door ports |
| Power | 110–240 VAC 50/60 Hz; 3 A @ 240 VAC / 6 A @ 110 VAC |
| Weight Dimensions | HG2-S 13 kg, 45×41×21 cm; HG2-XL 37 kg, 80×62×41 cm |
| Special Features | AutoCal / Remote API / range extensions optional; UV water sterilisation; heated sample loop |