Michell Instruments

Michell Instruments XTP601 Oxygen Analyzer

SKU · 42633 · XTP601 · Бренд: Michell Instruments
В наличии
Цена по запросу
Без онлайн-оплаты · строки в RFQ BOM
Скачать datasheet PDF

RFQ только BOM. Отправьте для инженерного КП.

Ключевые параметры

Accuracy
< ±1 % of range or ±0.02 % O₂ (whichever greater); suppressed-zero class ±0.2 % O₂ or ±1 % of range
Wetted parts
316L stainless steel, borosilicate glass, platinum, O-rings (Viton/Silicone/Ekraz options)
Hazardous area
ATEX/IECEx etc. Ex variants; SIL2 capable (OEM literature)
Technology
Thermo-paramagnetic oxygen measurement
Measurement Ranges
From 0…0.5 % to 0…50 % O₂ and suppressed-zero 20/80/90…100 % O₂ (factory calibrated to ordered range)
Repeatability
±0.2 % of range or ±0.02 % O₂
Скачать datasheet PDF
ПараметрЗначение
Accuracy< ±1 % of range or ±0.02 % O₂ (whichever greater); suppressed-zero class ±0.2 % O₂ or ±1 % of range
Wetted parts316L stainless steel, borosilicate glass, platinum, O-rings (Viton/Silicone/Ekraz options)
Hazardous areaATEX/IECEx etc. Ex variants; SIL2 capable (OEM literature)
TechnologyThermo-paramagnetic oxygen measurement
Measurement RangesFrom 0…0.5 % to 0…50 % O₂ and suppressed-zero 20/80/90…100 % O₂ (factory calibrated to ordered range)
Repeatability±0.2 % of range or ±0.02 % O₂
Sample Flow Gp100…500 ml/min (0.25…1.0 scfh) general purpose
Sample Flow Ex270…330 ml/min (0.57…0.7 scfh) Ex version
Sample Pressure0.75…2 BarA (calibrate at operating pressure)
Sample GasNon-condensing sample with particles <3 μm
OutputsDual 4–20 mA; Modbus RTU (RS485); concentration alarms
Ingress ProtectionIP66 / NEMA 4X (XTP601); IP55 (XTP501 indoor)
Dimensions Xtp601234 × 234 × 172 mm
Sample System OptionConfigurable Michell PS601 sample conditioning systems available for process integration