Panametrics
Panametrics air.IQ Sample System
Цена по запросу
Без онлайн-оплаты · строки в RFQ BOM
Скачать datasheet PDF
RFQ только BOM. Отправьте для инженерного КП.
Ключевые параметры
Measuring range
–110 to +20 °C dew/frost point (system); dew.IQ overall –110 to +60 °C capability; standard –80 to +20 °C with data to –110 °C
Accuracy
Calibrated accuracy at 77 °F (25 °C): ±2 °C above –100 °C; ±3 °C below –100 °C dew/frost point
Output
Isolated 4–20 mA / 0–20 mA / 0–2 V switch-selectable; two SPDT alarm relays; fail-safe fault relay
Supply
100–240 VAC 50/60 Hz or 24 VDC option
Process connection
1/4 in or 6 mm fittings (variant-dependent)
Display
128 × 64 matrix LCD (dew point °C/°F, ppmv with constant pressure input)
| Параметр | Значение |
|---|---|
| Measuring range | –110 to +20 °C dew/frost point (system); dew.IQ overall –110 to +60 °C capability; standard –80 to +20 °C with data to –110 °C |
| Accuracy | Calibrated accuracy at 77 °F (25 °C): ±2 °C above –100 °C; ±3 °C below –100 °C dew/frost point |
| Output | Isolated 4–20 mA / 0–20 mA / 0–2 V switch-selectable; two SPDT alarm relays; fail-safe fault relay |
| Supply | 100–240 VAC 50/60 Hz or 24 VDC option |
| Process connection | 1/4 in or 6 mm fittings (variant-dependent) |
| Display | 128 × 64 matrix LCD (dew point °C/°F, ppmv with constant pressure input) |
| Approvals | EMC 2004/108/EC; LVD 2006/95/EC; IP66 EN60529; NEMA 4X UL508A |
| Sensor type | Thin-film aluminum oxide (IQ.probe) with stainless sensor shield |
| Repeatability | ±0.2 °C above –100 °C; ±0.5 °C below –100 °C |
| Process Pressure | 0 to 200 psig (0 to 1,379 kPa) sample inlet |
| Sample Temperature | 0 to +50 °C (32 to 122 °F) |
| Enclosure | Wall-mounted IP66 / NEMA 4X stainless steel with clear window; IK10 impact |
| Sample Gases | Inert gases such as air, nitrogen, or argon (safe-area) |
| Data Logging | 32 GB MicroSD capacity (4 GB card included) |
| Warm Up | Meets specified accuracy within three minutes |